Lines of code count not available
Done
I don't think we need this.
Status should only be TRUE if msg accepted and signal sent. Move into if statement above.
Remove blank line.
Use alarmStatus structure to check whether alarm resume is allowed. proprs only looks at the passed alarm.
This should not be necessary. Already included by AlarmMgmt.h which includes TDCommon.h which includes Common.h which includes AlarmMgmtTD.h.
Dialysate temperature alarms should be TRUE for auto resume.
Fixed.
When Systems gives me Leahi terms, they will be very small fractions.
You're right. Old comment from Denali. Fixed.
Came from Denali, so I believe this is what they wanted. The nearly 5 minutes is the one the user sees.