Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-11543: Unable proceed to the Confirm Treatment when "08-3251-9" Acid Concentrate is selected

The HD FW updated to accept up to 4 acid concentration type index.

Below is the UI DataList.conf for acid concentrate.

[Acid Concentrate]

Fres. Naturalyte

08-1251-1

08-2251-0

08-3251-9

Finishing up the story for firmware

DEN-10607: Implementation [Screen]

DEN-10609: Implementation [Treatment Log]

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Fixed DG_RESERVOIR to DG_RESERVOIRS

Merge branch 'DEN-10153_Treatment_Log_Prescribed_UF' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-10247: UI Acknow timeouts on HD side, due to the delay of almost 2 seconds.

- Added another timeout of 5000 ms ( same value as UI_COMM_TIMEOUT_IN_MS ) only for the INIT state in which UI has a delayed response.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 57 more files in changeset.
DEN-9906: Addressed code review feedback.

DEN-9906: Fixed doxygen comments in BloodFlow module.

DEN-9906 updated the code to support events

  1. … 4 more files in changeset.
DEN-9906: Fixed merge conflicts from staging.

  1. … 47 more files in changeset.
DEN-9906: Added new volume checks to saline bolus, rinseback and blood prime deliveries.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-9480 addressed the code review comments

DEN-9480 addressed the code review comments

DEN-9906: Removed obsolete build switch for old FMD sensor.

DEN-9480 addressed the code review comments

DEN-9906: Added build switch for disabling UI comm timeout alarm.

DEN-9906: Fixed misc TBDs in comments.

DEN-9480 added stop RTC command

DEN-9906: Added more precision to dialysate flow filtering.

DEN-9906: Fixed bug w/ dialysate flow freshness alarm logic.

DEN-9906: Minor changes to dialysate flow freshness alarm code.

DEN-9906 Fixed compiling errors