Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-11750 addressed the code review comments

DEN-11750 addressed the code review comments

DEN-11750 addressed the code review comments

    • -10
    • +10
    ./Controllers/TemperatureSensors.c
DEN-11750 addressed the code review comments

DEN-11750 added a build switch for heat disinfect

DEN-11750 updated the comments

DEN-12232 fixed the fans monitor

DEN-11750 addressed the code review comments

DEN-11750 updated the drain exit algorithm

DEN-11750, DEN-12221 updated the drain exit algorithm

DEN-11750 updated the concentrate pumps

DEN-12225, DEN-12221 added code to update the UV reactors health and to calculate the target flow change based on the target drain pump RPM

DEN-12225 updated the UV reactors

DEN-11750 updated the code

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' of ssh://dvm-linux02:7999/dg/dgfirmware into DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up

DEN-11750: Fixed casting error in ConcentratePumps.c.

DEN-11750 addressed the bugs that Code Clinic reported

DEN-11750 added more comments

DEN-11750 worked on the RO pump duty cycle feedback

DEM-11750 updated the code

DEN-11750: Fixed drain pump RPM out of range alarm.

DEN-11750 updated the drain pump

Updated code to store the raw conductivity data from sensors CD1 and CD2

DEN-11750 updated the drain pump RPM to make sure the pump's RPM is only checked if the pumpis running

DEN-11250 Updated conductivity sensors manufacture information

    • -1
    • +15
    ./Controllers/ConductivitySensors.h
DEN-11750 Fixed temperature compensated conductivity error and added code to publish raw conductivity data. Updated FW code per peer review comments

    • -90
    • +9
    ./Controllers/ConductivitySensors.c
DEN-11750 enabled the load cells tare alarm

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' of ssh://dvm-linux02:7999/dg/dgfirmware into DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up

DEN-11750: Renamed msg ack ID for consistency with HD code.

DEN-11750 Fixed temperature compensated conductivity error and added code to publish raw conductivity data

    • -1
    • +91
    ./Controllers/ConductivitySensors.c