Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
DEN-15183: Fix for DEN-15198.

DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 30 more files in changeset.
DEN-14999 updated the alarms table

  1. … 20 more files in changeset.
DEN-14999 updated the code based on the alarms table

  1. … 2 more files in changeset.
DEN-14999-2: ignore venous occlusion alarm if venous pressure alarms disabled.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-14999: merged staging

  1. … 2 more files in changeset.
DEN-14999: updates to pressure POST test and 24V range alarms.

  1. … 1 more file in changeset.
Fixed merge conflict from staging.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
DEN-14906: fixed merge conflicts from staging.

  1. … 11 more files in changeset.
DEN-14906: Fixed arterial high pressure check while in adjunct treatment states.

DEN-14906: Addressed code review comment.

DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 6 more files in changeset.
DEN-14999: Reduced BP occlusion clear threshold from 5500 to 5000 above baseline.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
DEN-13154: Removed obsolete pressure data from broadcast. Tightened no cartridge pressure range for PBA and PBo.

  1. … 1 more file in changeset.
DEN-13154: Ensure alarm system has control of alarm lamp in fault mode in case fault occurred before lamp test completed. Do not request DG active alarms on startup if DG is not communicating.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Merge branch 'staging' into DEN-14589-DIAL-211

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 46 more files in changeset.
DEN-14549: Fixed merge from staging merge conflicts.

  1. … 17 more files in changeset.
DEN-14556: updated alarm pressure occlusion alarm to 3s. HD now sends events when occlusion baseline is set. Alarm id 142 and 277 now send baseline occlusion value and trigger value.

DEN-14549: Added alarm threshold to occlusion alarm data for logging. Only check for sufficient Heparin when in pre-tx mode. Addressed code review comments.

  1. … 2 more files in changeset.
DEN-14549: Update to arterial/venous pressure limit alarm logic. Update measured UF volume in all dialysis states. Remove dialysis and UF start states.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14549: Fix art/ven pressure alarm check logic - apply user set limits in dialysis state (if not saline bolus) or treatment stop state (if BP running). Require door/latch to be closed in patient connection state. Remove escalation for prime complete alarms. Handle prime complete alarms in pre-tx recirculate state. Update prime complete alarm properties.

  1. … 6 more files in changeset.
DEN-14011: created handleDrySelfTestStartState. code review comments.

  1. … 1 more file in changeset.