WatchdogMgmt.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
RESOLVED IN CODE WALKTHGROUGH

RESOLVED IN CODE WALKTHGROUGH

We did not check for the upper limit in the past, does this add new rejection scenarios

We did not check for the upper limit in the past, does this add new rejection scenarios

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add minimum check here too.

Add minimum check here too.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Do we need to add check for minimum treatment duration?

Do we need to add check for minimum treatment duration?

Cast constant as (F32) as before.

Cast constant as (F32) as before.

Done

Done

dialVolume is a volume (not a flow). Use constant (as before) for dialVolume check.

dialVolume is a volume (not a flow). Use constant (as before) for dialVolume check.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

This alarm should be restored.

This alarm should be restored.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

I check these against result since it is result &= ().

I check these against result since it is result &= ().

Done

Done

Remove blank line.

Remove blank line.

All of these ternaries should say ... ? result : FALSE );

All of these ternaries should say ... ? result : FALSE );

HD-DEN-16693_Institutional Settings Implementation
HD-DEN-16693_Institutional Settings Implementation
HD-DEN-11750_DG Dev Dialysate Temperature Control Tune Up
HD-DEN-11750_DG Dev Dialysate Temperature Control Tune Up
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH