DEN-14262: Added a/c power loss flag to alarm info broadcast.

  1. … 1 more file in changeset.
DEN-14262: Force re-home in no cartridge self-tests state. Addressed DEN-14171.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming' into staging

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Addressed code review comments.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Addressed paren typo from code review

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Additional Code review comments addressed.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Moved Alarms blocked timer variable, per review.

DEN-15183: Fixed compiler warning.

DEN-15183: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Comments updated.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Updated allowed alarms during PF

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Switched to CPLD power loss function, rather than alarm.

DEN-15183: Prevent alarm user options while AC power is out.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Moved block to choke point in activateAlarm2Data. Bock all alarms, rather than sub-rank 10 and above.

DEN-15183: Implemented updates to treatment recirculation workflow to include states for instructing user to disconnect and reconnected patient. Updated scalar for alarm audio current high gain signal. Updated alarm audio POST.

  1. … 2 more files in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adding an alarm blocking state to prevent cascade of errors during power fail.

DEN-14183: Addressed DEN-15190 by allowing rinseback option regardless of whether alarm condition is cleared.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 19 more files in changeset.
DEN-15105: Started adding pre-tx patient connection sub-states.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15105: Fix for DEN-15154 - saline bolus issue.

  1. … 3 more files in changeset.
DEN-15105: Removed alarm 227. Fixed bug DEN-15102.

  1. … 2 more files in changeset.
DEN-15110 Use 800ml/min reservoir fill rate for all dialysate flow rates.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 27 more files in changeset.
Merge branch 'staging' into DEN-14999-alarm-table-updates-part-2

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14999 updated the alarms table

  1. … 20 more files in changeset.