Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DEN-17299_Release 22 P1B Staging
HD-DEN-17299_Release 22 P1B Staging
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Done

Done

This is not really a rej ratio - it's a maximum reg ratio. We should make that clear in naming #defines, field names, etc...

This is not really a rej ratio - it's a maximum reg ratio. We should make that clear in naming #defines, field names, etc...

HD-DENBUG-318_Rejection Ratio Setting Feature Implementation IN Manufacturing Configuration For Treatment Staging
HD-DENBUG-318_Rejection Ratio Setting Feature Implementation IN Manufacturing Configuration For Treatment Staging
HD-DENBUG-298_User Unable TO Deliver Saline P1B Staging
HD-DENBUG-298_User Unable TO Deliver Saline P1B Staging
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Bad practice to have two calls to same function here. Possible that 2nd call will return different result. Should call function once to assign value to a local var prior to if statement and then re...

Bad practice to have two calls to same function here. Possible that 2nd call will return different result.
Should call function once to assign value to a local var prior to if statement and then reference the local var in the if statement.