Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-16372: Addressed DEN-16109 - increased alarm threshold for LC drift from 10 to 20g.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-16456-dpo-does-not-home-in-startup-or-end-treatment' into staging

DEN-16456 Fix DPO homing speed.

DENBUG-78: increased alarm 110 persistence from 3 to 6 seconds to better handle slow dispense rates.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 17 more files in changeset.
Merged staging to DEN-15792.

  1. … 7 more files in changeset.
DEN-15792: Fixed some comments.

DEN-16456 Use basic PWM calculation for Homing DPO.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-16385-treatment-log-event-ids-for-heparin-pause-are-not-notated-correctly' into staging

DEN-16385 Fix pause message when requested by user. Move HeparinDeliveryState set to after calling StopSyringePump() which needs it for pause message to be sent.

DEN-16382 added code to not send the DG fill command if the empty bottle alarms are active

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-16385-treatment-log-event-ids-for-heparin-pause-are-not-notated-correctly' into staging

DEN-16385 Remove redundant event data messages. Send only when pump started/stopped.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

STT-2824 changed the location of the blood leak setpoint failure event

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16385 Send heparin stop/pause event when syringe pump stopped.

DEN-16408: Updated DPi and DPo modules to increase speed range to 0..4000 RPM in order to allow pumps to reach 600 mL/min rate consistently.

HD-DEN-16314-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Removed unused elements of Reservoir data.

DEN-16385 Fix heparin pause command ID.

DEN-16365 Enable venous occlusion alarm in treatment recirculation mode. Other pressure alarms disabled.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16200-2-merge in to staging

DEN-16200-2-syringe-pump-stuck-in-priming-state