UI-LEAH-1924_Leahi/Features/Leah 1924 UI AM S4 Translation
DEN-15493 : Usability: UI, FW: rinseback flow rate

- Updated the rinseback ranges regarding the jira ticket description.

- 100 to 300, 200 as default.

DEN-15319: Usability: Acid Bicarb Prescription Updates

DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- confirmed the ranges with Blaine and updated the ranges and PRS,SRSs

DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

DEN-15407: UI: Add RO Permeate Sample to cleaning mode screen + messaging

DEN-15316: Usability: UI: Remove Extra Button for Disinfection

  1. … 2 more files in changeset.
DEN-15062 UI: update venous pressure alarm limits (venous pressure high max = 300 - was: 400)

- Changed the default as well for that range.

DEN-15062 UI: update venous pressure alarm limits (venous pressure high max = 300 - was: 400)

Re-committed the delete files/folders

  1. … 63 more files in changeset.
Bamboo Commit: Updated the unhandled messages config file

  1. … 62 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

removed spaces.

removed spaces.

to align arterial(8 characters) and venous(6 characters) vertically, and to make it easier to compare the variables, to avoid missing a parameter.

to align arterial(8 characters) and venous(6 characters) vertically, and to make it easier to compare the variables, to avoid missing a parameter.

Why is there 2 spaces between . and field name?

Why is there 2 spaces between . and field name?

DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- changed the low/high of the monitoring ranges to min/max.

DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- Added all the Treatment parameters to the DataList.conf so the UI can read the parameters from the settings instead of hard coding them.

UI-DEN-13962_UI BN S82
UI-DEN-13962_UI BN S82
DEN-13962: UI: Update pressure limits to match treatment params and UI screen

- Added all the Treatment Parameters to the DataList.conf to give system to update independently.

DEN-13489 IMP: General Implementation

DEN-13035: Update alarm body messages

- date/time format merged into the Settings/System.conf, and removed.

- the Alarms/Alarms.conf has been generated from the Alarms.xlsx by a python script, pushed into scripts repo.

  1. … 2 more files in changeset.
DEN-12639: Change Request: Add Larger Dialyzer

- Added bigger dialyzer

DEN-12512: INT: Support/Meetings/Planning [ Removed incorrectly added extra duplicate line.]

DEN-12512: INT: Support/Meetings/Planning

DEN-09712: Date format between system date/time and BP/HP "Last Read" is not consistent

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

DEN-12512 INT: Support/Meetings/Planning/Development

- Renamed the TreatmentParameters folder to the Configurations to give it a more common name.