Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
Relocated app image, Created new modules for update functions, First implementation of update...
Relocated app image, Created new modules for update functions, First implementation of update...
Relocated app image, Created new modules for update functions, First implementation of update...
Relocated app image, Created new modules for update functions, First implementation of update...
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DEN-17299_Release 22 P1B Staging
HD-DEN-17299_Release 22 P1B Staging