DEN-9906: Fixed merge conflicts from staging.

  1. … 61 more files in changeset.
DEN-9906: Added test code to calculate and report measured UF rate based on change in UF volume.

DEN-9906: Optimized HD broadcast messaging.

  1. … 37 more files in changeset.
DEN-9906: Addressed DEN-8719. Finished sending sub-mode change events. Minor changes per DEN-10202.

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 111 more files in changeset.
DEN-9906: Fixed target rinseback volume published after rinseback stopped prematurely. Allow alarm response resume if in pause state of Tx End sub-mode.

  1. … 3 more files in changeset.
DEN-9906: Fixed poor UF control start.

DEN-9906: Added tentative blood flow estimation logic for testing. Added HD power-off function. Power off HD if no AC power on startup. Added t/o to rinseback stop state. Fixed valve commands in post-treatment mode.

  1. … 15 more files in changeset.
DEN-8886: Addressed code review comments.

  1. … 7 more files in changeset.
DEN-8886: Fixed typo in DPo speed check.

DEN-8886: Local merge from staging.

  1. … 36 more files in changeset.
DEN-8886: Finished adding overrides for mode broadcast intervals.

  1. … 22 more files in changeset.
DEN-8886: Added flow range checks. Added pressure range checks.

  1. … 7 more files in changeset.
DEN-8886: More HD pump tuning based on data from lab testing.

  1. … 3 more files in changeset.
DEN-8886: Tuned pump controls per lab test data.

  1. … 4 more files in changeset.
DEN-8886: Updated pump control.

  1. … 7 more files in changeset.
Added build switch to allow higher pump set speeds.

  1. … 3 more files in changeset.
DEN-8103: Fixed merge issue

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-8103-dg-hd_dev-alarm-design

  1. … 25 more files in changeset.
DEN-7605: Added multiplier for initial DPo PWM estimate to accomodate observed delta.

DEN-7605: Added pump direction error counter check and alarm. Added build switch to allow change to treatment duration to as low as 1 minute.

  1. … 5 more files in changeset.
DEN-8103: Fixed build warnings

  1. … 13 more files in changeset.
DEN-7605: Updated peristaltic pump speed over PWM DC range for V3.

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-7091-pre-treatment-mode

  1. … 20 more files in changeset.
DEN-7091: Combined self-tests states and removed current min check when pumps running

  1. … 3 more files in changeset.
DEN-6372: merged from staging.

  1. … 13 more files in changeset.
DEN-6372: UF control change.

DEN-6372: Moved blood sitting timeout logic from alarm escalation to treatment stop sub-mode.

  1. … 14 more files in changeset.
DEN-5887: Fixed merge conflicts from staging.

  1. … 13 more files in changeset.
DEN-5887: No adjustment to DPo initial PWM at this time. Restored P and I control terms to original settings.