Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DEN-17299_Release 22 P1B Staging
HD-DEN-17299_Release 22 P1B Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DENBUG-318 changed the RO rejection ration in institutional to percent and min as in integer

  1. … 1 more file in changeset.
DENBUG-318 added a max limit #define for the min inlet water cond alarm. Updated the HD institutional record structure that is sent to the UI

  1. … 1 more file in changeset.
DENBUG-318 updated the code to include all the DG institutional values upon requesting from DG in HD

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Done

Done

DENBUG-320 added the word max to the RO rejection ratio

  1. … 2 more files in changeset.
This is not really a rej ratio - it's a maximum reg ratio. We should make that clear in naming #defines, field names, etc...

This is not really a rej ratio - it's a maximum reg ratio. We should make that clear in naming #defines, field names, etc...

DENBUG-318 added code to range check the RO rejection ratio in HD institutional record and send it to DG upon request from DG

  1. … 1 more file in changeset.
HD-DENBUG-318_Rejection Ratio Setting Feature Implementation IN Manufacturing Configuration For Treatment Staging
HD-DENBUG-318_Rejection Ratio Setting Feature Implementation IN Manufacturing Configuration For Treatment Staging