Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DENBUG-104: Fixed errors in bug fix.

  1. … 1 more file in changeset.
DENBUG-104: restored max dialysate pump speed to 4000 RPM. Apply reduced lower rotor speed tolerance only to commanded speeds > 3200 RPM.

  1. … 1 more file in changeset.
DENBUG-104: Increased max dialysate pump speed from 3400 to 3600 RPM. Reduced lower rotor speed tolerance from -5 RPM or -15% (whichever is bigger) to -60%.

  1. … 1 more file in changeset.
DENBUG-104: Allow for dialin pwm set cmd to go to full speed.

  1. … 1 more file in changeset.
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
HD-DENBUG-104_Alarm 38 HD Fault Dpi Rotor T
DENBUG-104: Cap dialysate pump speed to 3,400 RPM to prevent rotor speed errors.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

DEN-16518 code review changes

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
DEN-16518-alarm-41-triggered-on-dvt-014-at-the-beginning-of-treatment

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.