HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
DENBUG-328 added an air trap event

DENBUG-328 updated the alarm title of alarm 215

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-328

DENBUG-328 updated the text of alarm 215

RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

DENBUG-328 updated the UI text of alarm 215

DENBUG-327 updated the DG FPGA compatibility revision

DENBUG-327 added a DG event for the baro sensor mfg CRC check

Changed the rank of alarm 215 to 200

Changed the alarm properties of alarm 215. It is dialysate recirculation

DENBUG-328 updated the softawre compatibility

DENBUG-328 updated the air trap upper level fluid timeout alarm properties

DENBUG-328 added a new alarm for the air trap upper level detecting fluid for too long

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging