firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-334 reverted the heat disinfect reservoir transfter target flow back 0.5 L/min

updated the code to initialize the variables in between the treatments without a need to power cycle the unit

DG-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
DG-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DENBUG-327_Baro Sensor Data Refresh Staging
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
    • -2
    • +2
    ./App/Controllers/TemperatureSensors.h
DENBUG-327 updated the VectorCAST test cases

DENBUG-327 updated the baro sensor mfg CRC check

    • -5
    • +7
    ./App/Controllers/TemperatureSensors.c
Done

Done

DENBUG-327 added code to make the baro sensor MFG CRC as LSB

Do we need to shift CRC bits to lsb position?

Do we need to shift CRC bits to lsb position?

DENBUG-327 updated the baro sensor code

    • -16
    • +28
    ./App/Controllers/TemperatureSensors.c
    • -0
    • +1
    ./App/Controllers/TemperatureSensors.h
DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DEN-17299_Release 22 P1B Staging
DG-DEN-17299_Release 22 P1B Staging
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -2
    • +2
    ./App/Controllers/ConductivitySensors.h