updated the code to initialize the variables in between the treatments without a need to power cycle the unit

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DENBUG-327 updated the baro sensor mfg CRC check

  1. … 1 more file in changeset.
DENBUG-327 updated the baro sensor code

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DENBUG-318 changed the min inlet water conductivity to recovery offset to 10 uS/cm. Updated the vectorcast test cases

DENBUG-318 removed the inlet recovery conductivity alarm #define

DENBUG-318 changed the inlet water conductivity low alarm recovery

DENBUG-318 added a new event for min inlet water conductivity alarm from HD institutional record

  1. … 1 more file in changeset.
DENBUG-318 updated the code to receive all the DG institutional record items. Added code to use the min inlet water conductivity alarm from the institutional record

  1. … 5 more files in changeset.
DENBUG-322 added code to set CPi and CPo to 20 uS/cm if the values are less than 20 uS/cm

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-17263 addressed the code review comments

DENBUG-314 added more comment

DEN-17263 fixed the code to check the direction of the concentrate pumps and then decide based on that whether the measured speed values being negative or positive is correct or not

DEN-17263 added code to check the concentrate pumps speed direction

DENBUG-307 added code to broadcast the concentrate pumps speed the way they are regardless of the direction of the pump

DENBUG-307 made the concentrate pumps speed check all absolute to be able to check both positive and negative targets and measured speeds

DEN-17263 change the concentrate pumps speed check to not use fabs for the measured values. The target speed is checked in fabs to cover the negative target speeds

DENBUG-309 added mode flush to the 5% tolerance of the concentrate pumps speed check

DENBUG-309 changed the concnetrate pumps speed monitor to 5% in heat disinfect

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 86 more files in changeset.
DENBUG-289 added more comments

DEN-17135: added code to check the concentrate pumps calibration reocrd live. DEN-17146: fixed the active cool mode

  1. … 2 more files in changeset.
DEN-17135 updated the concentrate pumps calibration

  1. … 1 more file in changeset.
DEN-17135 added the calibration record to the acid and bicarb concentrate pumps

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-254 changed the steps that the baro temperature is calculated

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-232 fixed the tare code in the service mode