PresOccl.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-334 updated the code to initialize the variables in between the treatments without a need to power cycle the unit

  1. … 9 more files in changeset.
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
HD-DENBUG-249_Pressure Window Starting Values P1 Staging
HD-DENBUG-249_Pressure Window Starting Values P1 Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done

Done

Done

Done

Done

Done

Yes, when we are not in treatment mode, the state machine sets the state to "pressure_limits_state_off" at start of this function ( line : 673). hence mode check is not necessary.

Yes, when we are not in treatment mode, the state machine sets the state to "pressure_limits_state_off" at start of this function ( line : 673). hence mode check is not necessary.

Done.

Done.

Done.

Done.

I think we just want to publish the baseline here (not +1000).

I think we just want to publish the baseline here (not +1000).

Yes, if you cast either value in division as a F32, the result will be F32. This is all we need as F32 (the .data). The samples and total can stay U32. You might want to cast the total as F32 as we...

Yes, if you cast either value in division as a F32, the result will be F32. This is all we need as F32 (the .data). The samples and total can stay U32.
You might want to cast the total as F32 as well for consistency.

F32 not needed here.

F32 not needed here.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Casting not necessary.

Casting not necessary.