Verify this is not a problem. It resets some UF stuff. But we are calling this every time we transition to Tx: Stop .

Verify this is not a problem. It resets some UF stuff. But we are calling this every time we transition to Tx: Stop .

DENBUG-335: air trap fill refinements.

  1. … 4 more files in changeset.
DENBUG-335: First cut at new air trap fill approach with closing of VBV and slowing of BP.

  1. … 7 more files in changeset.
HD-DENBUG-335_Air Trap Fill Design Improvem
HD-DENBUG-335_Air Trap Fill Design Improvem
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle P1B Staging
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle P1B Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 104 more files in changeset.
DENBUG-334 updated the code to initialize the variables in between the treatments without a need to power cycle the unit

  1. … 2 more files in changeset.
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-328 updated the air trap event and the VectorCAST test cases

DENBUG-328 changed the get raw air trap function to get air trap function for alarm 215. Added air trap state change events

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH