DENBUG-335 implemented the changes from staging

  1. … 1 more file in changeset.
DG-DENBUG-335_Air Trap Fill Design Improvements
DG-DENBUG-335_Air Trap Fill Design Improvements
HD-DENBUG-335_Air Trap Fill Design Improvem
HD-DENBUG-335_Air Trap Fill Design Improvem
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Done

Done

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

The comment was left here to make sure the next developer does not ask the question that the blood pump was not set here. It is a note to remind the developer that the ramp is set below.

The comment was left here to make sure the next developer does not ask the question that the blood pump was not set here. It is a note to remind the developer that the ramp is set below.

Can we move the comment where required?

Can we move the comment where required?

Dara, work with Nick to get the right text for this new alarm.

Dara, work with Nick to get the right text for this new alarm.

We changed the order of ORs. But the parenthesis match otherwise, it would not compile.

We changed the order of ORs. But the parenthesis match otherwise, it would not compile.

Done

Done

Done

Done

No this comment is explaining what will happen at the bottom.

No this comment is explaining what will happen at the bottom.

Need to remove comment? or missing implementation?

Need to remove comment? or missing implementation?

The alarm data doesn't sound correct.

The alarm data doesn't sound correct.

Remove extra blank line

Remove extra blank line

This alarm description is more similar to alarm 215, including title, description, trigger conditions?

This alarm description is more similar to alarm 215, including title, description, trigger conditions?

This doesn't look right. Check parenthesis.

This doesn't look right. Check parenthesis.

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
HD-DENBUG-298_User Unable TO Deliver Saline Bolus During Alarm Conditions Where The Blood Pump IS Recirculating
DENBUG-328 added code from staging

  1. … 3 more files in changeset.
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi