You are calling setAirPumpState in if and else. Consider just calling once after the if/else.

You are calling setAirPumpState in if and else. Consider just calling once after the if/else.

Add space between if and (

Add space between if and (

Fixed

Fixed

LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
LEAHI-TD-FIRMWARE-LDT-3600_Request for TD Test Config for Testing without Tubing Set on TD
Comment to right no longer makes sense.

Comment to right no longer makes sense.

Should this event be TD_EVENT_AIR_TRAP_FILL?

Should this event be TD_EVENT_AIR_TRAP_FILL?

LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
fixed in latest code

fixed in latest code

implemented

implemented

fixed

fixed

I think this would be a TD s/w fault (alarm already define). Just need a new TD s/w fault ID in AlarmMgmtSWFaults.h.

I think this would be a TD s/w fault (alarm already define). Just need a new TD s/w fault ID in AlarmMgmtSWFaults.h.

This structure doesn't seem necessary - it's just a middle man. Why not just pass the #defines directly into the quadratic macro?

This structure doesn't seem necessary - it's just a middle man. Why not just pass the #defines directly into the quadratic macro?

Fixed

Fixed

removed from here

removed from here

fixed

fixed

We need new alarm . That should be included in SRS right ? or shall i use SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_SOFTWARE_FAULT) ?

We need new alarm . That should be included in SRS right ? or shall i use SET_ALARM_WITH_2_U32_DATA( ALARM_ID_TD_SOFTWARE_FAULT) ?

Add blank line after declarations.

Add blank line after declarations.

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Alarm should be handled inside of setAirPumpState().

Add blank line after declarations.

Add blank line after declarations.

I think alarm should be handled inside the setAirPumpState() function.

I think alarm should be handled inside the setAirPumpState() function.

Should we have alarm here?

Should we have alarm here?

duty misspelled.

duty misspelled.

Can we remove inclusion of fpga here? It shouldn't be needed.

Can we remove inclusion of fpga here? It shouldn't be needed.