It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD...

It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD or HDF type). Then when we get this response msg from UI, we will compare the type to the user selected modality (from TxParams) to see if it matches.

Add doxygen comments to the right of each structure field.

Add doxygen comments to the right of each structure field.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

Add blank line before return statement.

Add blank line before return statement.

Keep this blank line above group header.

Keep this blank line above group header.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
LDT-1886: Update event related code

We can rename priorState to previousInstallState. Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

We can rename priorState to previousInstallState.
Also, it should be declared as a static variable to be able to retain previousState value when the control goes out of the function.

LDT-1886: Add send event

These alarms are in the wrong place. Their enums are 187..199. So they should be at the very end of this table, not here near the beginning.

These alarms are in the wrong place. Their enums are 187..199. So they should be at the very end of this table, not here near the beginning.

Remove this blank line.

Remove this blank line.

There is still merge conflict? Please resolve merge conflict and build the code in your local before pushing.

There is still merge conflict?
Please resolve merge conflict and build the code in your local before pushing.

This merge conflict is still unresolved.

This merge conflict is still unresolved.

leak in the comment section.

leak in the comment section.

Where is alarm 191..203?

Where is alarm 191..203?

LDT-1886: Update doxygen stuff

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

please add bpLeftHomeTimerCounter to inputs and outputs.

LDT-1886: Refactor tube set jam logic

  1. … 1 more file in changeset.
LDT-1886: Update alarm name blood tube set failure

LDT-1886: Remove backOffAfterRetract variable and update brief

  1. … 1 more file in changeset.
I missed updating this earlier, I’ll push the updated code.

I missed updating this earlier, I’ll push the updated code.

Can't have 3 values. Maximum 2 values allowed.

Can't have 3 values. Maximum 2 values allowed.

Removedd

Removedd

added

added