leahi-fwcommon

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Yes, updated

Yes, updated

Removed

Removed

Remove blank line.

Remove blank line.

Looks like this reverted back.

Looks like this reverted back.

Fixed it

Fixed it

Updated

Updated

Fixed

Fixed

Remove extra blank line.

Remove extra blank line.

It is your comment, so please resolve it.

It is your comment, so please resolve it.

It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

For these functions that set specific valves, we should specify which (e.g. recovery valves or balancing chamber valves, or ...).

For these functions that set specific valves, we should specify which (e.g. recovery valves or balancing chamber valves, or ...).

Remove blank line.

Remove blank line.

It will be fixed in Balancing chamber re-implementation

It will be fixed in Balancing chamber re-implementation

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

fixed it

fixed it

fixed it

fixed it

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working ) //Increment counter for dry bicarb c...

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working )

//Increment counter for dry bicarb chamber fill
bicarbChamberPeriodicFillCounter += 1;
// Fill bicarb chamber once every 60secs.
if ( bicarbChamberPeriodicFillCounter >= BICARB_CHAMBER_PERIODIC_FILL_TIME )

Unknown macro: { setBicarbChamberFillRequested(); bicarbChamberPeriodicFillCounter = 0; }


The request is cleared here. for consistency I am moving to the end of the fill cycle

LEAHI-DD-FIRMWARE-LDT-3343_DD FW control stops IOFP manual control
LEAHI-DD-FIRMWARE-LDT-3343_DD FW control stops IOFP manual control
Mis-spelled "Stabilized".

Mis-spelled "Stabilized".

I think you can do this in 1 line of code using the RANGE macro in Common.h.

I think you can do this in 1 line of code using the RANGE macro in Common.h.

LEAHI-TD-FIRMWARE-LDT-1415_Pressures Monitoring - TD
LEAHI-TD-FIRMWARE-LDT-1415_Pressures Monitoring - TD
Yes, we do and this is not required here so, I removed

Yes, we do and this is not required here so, I removed

Shouldn't we be getting these defaults from the TxParams unit?

Shouldn't we be getting these defaults from the TxParams unit?