Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' into develop

  1. … 7 more files in changeset.
DEN-13460 updated the code

  1. … 12 more files in changeset.
Merged DEN-13460

  1. … 3 more files in changeset.
DEN-13460 updated the concentrate pump speed check and the increased the heaters on with no flow to 10 seconds

  1. … 6 more files in changeset.
DEN-13460 Heater feedback primary channel changed to read from FPGA, secondary from internal ADC replaces old primary name.

  1. … 6 more files in changeset.
Merged DEN-13460

  1. … 10 more files in changeset.
DEN-13460 updated the drivers with the DVT features

  1. … 9 more files in changeset.
Merge branch 'DEN-13427-sw_dev-sprint-75_mg' into develop

  1. … 22 more files in changeset.
Merge branch 'staging' into DEN-13427-sw_dev-sprint-75_mg

  1. … 3 more files in changeset.
DEN-13460 merged staging

  1. … 21 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 29 more files in changeset.
DEN-12931 addressed the code review comments

  1. … 4 more files in changeset.
DEN-12931 addressed the code review comments

  1. … 9 more files in changeset.
DEN-13460 updated the barometric sensor CRC check

  1. … 3 more files in changeset.
Merge branch 'DEN-13427-sw_dev-sprint-75_mg' into develop

DEN-13427: fixed warnings

Merge branch 'DEN-13427-sw_dev-sprint-75_mg' into develop

DEN-13427: Reworked DG Service mode handling

DEN-13460 updated the baro sensor and commented out the FPGA POST

  1. … 1 more file in changeset.
Merged DEN-13460

  1. … 5 more files in changeset.
DEN-13460 updated the code with the DVT changes

  1. … 7 more files in changeset.
DEN-13460: fixed merge conflicts.

DEN-13460: Handle request to re-send active alarms.

DEN-13460 Implemented FAULT LED and Audio control in AlarmMgmt.c.

Merge branch 'DEN-13427-sw_dev-sprint-75_mg' into develop

DEN-13427: Errors and warnings fixes

Complication Fix

DEN-13427: DG Service Mode deenergizes and can Fault

Service Mode deenergizes actuators when we enter the mode. Fault is raised in Service Mode if we get a submode out of range.

  1. … 2 more files in changeset.
DEN-13427: Message handler and response for DG's request to service mode

DEN-13460: If HD commands stop fill, clear bad fill flag in gen idle mode so we don't get stuck there.

Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' into develop

  1. … 2 more files in changeset.