Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 619 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 28 more files in changeset.
DENBUG-199: Alarm 280 (DG Fault: DRP Direction Fault) Triggered during Pre-Treatment Sequence

Added the Maxon RPM to the handled messages to be logged.

  1. … 6 more files in changeset.
DENBUG-163: High Venous Pressure Alarm 130 Triggers During Normal Treatment Operation

Added air trap raw lower level and air trap raw upper level to be logged and displayed on the diagnostics screen

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-16444: UI: UI Logs do not Capture TDi and TRo Moving Average Values

Updated the application to incorporate the two new float values in the DG temperature sensors payload.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
DENBUG-204 aligned the commented out code in the diagnostics qml

DENBUG-204 updated the diagnostics screens qml with the latest dialysate outlet pump broadcast changes

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
  1. … 1 more file in changeset.
DEN-15078 DRP Cur and HDrst Values Display Too Many Float Numbers

  1. … 2 more files in changeset.
Release feedback: Drain Pump currentA parameter precision

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 57 more files in changeset.
Merge branch 'DEN-14936-UI-BN-S93' into staging

  1. … 2 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 10 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 8 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

- MSG_ID 0x0B

- Added the keepAlive/UI-CheckIn to the list of AckRequired so if HD won't ack UI, the Alarm, "HD Communication Failure" will be displayed.

  1. … 6 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 19 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
DEN-13152 : Flow Rate Settings Display "--" or "0" When Treatment is Paused

- Messages 0x0500 and 0x0800 has been updated and the Prescribed Flow Rate is being used.

- The diagnostics screen updated.

- The treatment main screen updated.

- The treatment flow adjustment screen updated.

- the Flows on the commit has 10f39b63ce8a992280ab123f05a5eb44a27c121b has been separated which has to be reverted back since these two messages almost all the time have the same payloads.

  1. … 11 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 50 more files in changeset.
Merge branch 'DEN-14733-UI-BN-S90-2' into staging

  1. … 43 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
DEN-14391 , DEN-14393 : Qt 5.15.10 [QML run-time issues]

- Fixed all the QML warning and errors

- Enabled the Clang-Tidy, Clazy and cppcheck and fixed some.

- Removed unused images

- made the Denali POST logo transparent.

  1. … 56 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-12262: INT: Support and debugging

- Fixing the DG Valves/Temperature messages

- Fixing the CloudSync RegisterDeviceRequest

- Fixed/Cherry-Picked the Dialin-log conditions.

- Cleaned up the debug messages.

  1. … 20 more files in changeset.
DEN-12262: INT: FW Support (Reset Diagnostic Data)

- Reorganized the Diagnostics screen a little more.

  1. … 2 more files in changeset.