HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-328 updated the air trap event and the VectorCAST test cases

DENBUG-328 changed the get raw air trap function to get air trap function for alarm 215. Added air trap state change events

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

DENBUG-328 addressed the code review comments

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

DENBUG-328 updated the UI institutional record structure

DENBUG-328 updated the min venous wide window range code

DENBUG-328 updated the air trap upper level fluid detection code

DENBUG-328 added persistent alarm for the air trap upper level fluid detection

DENBUG-328 added min venous limit from the institutional record

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.