DIAL-262-fixes

Activity

DIALIN-DEN-15196-7 6

Keyboard shortcuts  
  • Summarize the review outcomes (optional)
     
    #permalink

    Details

    Warning: no files are visible, they have all been filtered.
    Participant Role Time Spent Comments Latest Comment
    Author & Moderator 16m 2 We need to display the last Saline event before we switch...
    Reviewer - Complete 5m 4 RESOLVED in CODE WALKTHROUGH.
    wbracken  (deleted user)
    Reviewer - 0% reviewed      
    Reviewer - 100% reviewed 4m    
    Reviewer - 95% reviewed 4m    
    jtaylor  (deleted user)
    Reviewer - 0% reviewed      
    Total   30m 6  
    #permalink

    Objectives

    • DIAL-262: fixed operation status event to have more accurate transitioning in between dialysis sub modes.
    • DIAL-262: removed old 4th state data when hd pauses in priming
    • DIAL-262: added additional event calls for UF and saline transitions.
    • DIAL-262: fixes
    • Merge branch 'staging' into DEN-15196-DIAL-262-fixes
    • DIAL-264: fixed blood prime not showing sub state in operation status event

    Branches in review

    #permalink

    Issues Raised From Comments

    Key Summary State Assignee
    #permalink

    General Comments

    There are no general comments on this review.
    /firmware/App/Modes/BloodPrime.c Changed
    /firmware/App/Modes/Dialysis.c Changed 3
    /firmware/App/Modes/OperationModes.c Changed
    /firmware/App/Modes/OperationModes.h Changed 3
    /firmware/App/Modes/Prime.c Changed

    Review updated: Reload | Ignore | Collapse

    You cannot reload the review while writing a comment.

    Log time