DD-LDT-1473

I am still keeping M4 valve for some more time, when there is a need to operate DD stand-alone system (Without FP, but with IO system). So M4 valve and couple of pressure sensors (M1 and M3) are st...

I am still keeping M4 valve for some more time, when there is a need to operate DD stand-alone system (Without FP, but with IO system). So M4 valve and couple of pressure sensors (M1 and M3) are still active in DD system for time being.

This is also breaking FP f/w. Why do you both have M4?

This is also breaking FP f/w. Why do you both have M4?

DD-LDT-1473_Ultrafiltration DD
DD-LDT-1473_Ultrafiltration DD
DD-LDT-1473_Ultrafiltration DD
DD-LDT-1473_Ultrafiltration DD
Thanks. Resolved

Thanks. Resolved

it is moved to DDDef.h, UI team is looking for this mapping for their (diagnostics page) development.

it is moved to DDDef.h, UI team is looking for this mapping for their (diagnostics page) development.

Where did this valve enum go?

Where did this valve enum go?

Done.

Done.

I have introduced two additional build flags in 'DDCommon.h' (ENABLE_ALARM_1 and ENABLE_ALARM_2) to currently stop triggering alarms due to hardware support etc., so. it should be safe for now. if ...

I have introduced two additional build flags in 'DDCommon.h' (ENABLE_ALARM_1 and ENABLE_ALARM_2) to currently stop triggering alarms due to hardware support etc., so. it should be safe for now. if still any specific alarms triggers, will try to disable those alarms.

2 blank lines above and below test support banners.

2 blank lines above and below test support banners.

Is this safe to enable now?

Is this safe to enable now?

Consider doing this in update compensated rate function (only once every 2 minutes when rate actually changes).

Consider doing this in update compensated rate function (only once every 2 minutes when rate actually changes).

Done.

Done.

Done.

Done.

Remove extra blank line.

Remove extra blank line.

Done

Done

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

With the updated FPGA logic for continuous flow rate, doesn't need this variable anymore.

Done.

Done.

Done.

Done.

Isn't there a common override function that does this?

Isn't there a common override function that does this?

Looks like you are calculating iteration in "mL/5sec", not "mL/sec".

Looks like you are calculating iteration in "mL/5sec", not "mL/sec".

Are we dosing UF in spurts like concentrates?

Are we dosing UF in spurts like concentrates?

Where did 5 second period come from?

Where did 5 second period come from?