HD-DEN-16512

HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Consider splitting this if and the setDialysisParams() function into two: 1) blood flow rate changed and 2) dialysate flow rate changed.

Consider splitting this if and the setDialysisParams() function into two: 1) blood flow rate changed and 2) dialysate flow rate changed.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Done.

Done.

5% seems like a reasonable max step. I don't think we want to go to ramp up/down states if > 5% change. I think we just want to limit the change to max 5% for a single control interval.

5% seems like a reasonable max step. I don't think we want to go to ramp up/down states if > 5% change. I think we just want to limit the change to max 5% for a single control interval.

Remove blank line.

Remove blank line.

Add space after if.

Add space after if.

HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate
HD-DEN-16512_Dpo Stops During Treatment After Reconfirming Dialysate Flow Rate