HD-DEN-16518

HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Updated code changes. Open loop still maintains the minimum PWM duty cycle as 10%.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

Do we want to have the 11% minimum here? I thought it was just for closed loop control.

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

This will cause open loop control mode to have a minimum of 11% instead of 10%. For open loop, pwmDC is assigned a value at line 295. Do we still need the #define for the 10% pwm lower limit?

HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment
HD-DEN-16518_Alarm 41 Triggered ON Dvt 014 AT The Beginning OF Treatment