HD-DEN-16778

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Once it goes into prime pause state -> say, user removes the clamping and resume the priming -> again this state executes to check the clamping condition. Now, venous pressure is in range and raw f...

Once it goes into prime pause state -> say, user removes the clamping and resume the priming -> again this state executes to check the clamping condition. Now, venous pressure is in range and raw flow rate more than minimum, we get into else state with last set flag ( primeDialysateOcclusionDetected) and change the priming state back to blood priming initial states.

Alarm above has stop property. So triggering alarm will take us out of this state and go to Prime Pause state. So this else will never happen I think.

Alarm above has stop property. So triggering alarm will take us out of this state and go to Prime Pause state. So this else will never happen I think.

This looks like a threshold. So name should start with MIN or MAX.

This looks like a threshold. So name should start with MIN or MAX.

HD-DEN-16778_System Does Not Check That Air
HD-DEN-16778_System Does Not Check That Air