DEN-13134: IMP: General Implementation for any bugs and any other implementations
DEN-12836: Treatment Report/log export modifications
- Improved the SDItem and renamed to SDCProgressItem.
- Added another separate progress circle component for the USB Driver [USBProgressItem]
- Adding Space check for USB as well to get the USBProgressCircle update.
- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)
- Improved USB button comonent
- Added a global byte size converter in Variable.qml
- Improved MainMenu component
- Added a notification for the TreatmentFlowBase.qml main component.
- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.
- Added an individual screen for the log export.
- Removed the Export Log and USB eject from the main settings option screen.
- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.
- Fixed the DGVersionResponse, log message.
- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)
- Added error handling to the Treatment Log/Report export
DEN-12726: Denali application crashing
- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.
DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version
DEN-13156: Settings screen menu should disappear on service mode
- Improved the MainStack.qml navigation rules.
- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.
- Improved the SettingsStack.qml navigation rules.
DEN-13128: KBM: Cloud Sync Integration
- Saving the CS log in UTC timezone.
DEN-13127: LTS: Development Support
- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.