Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-13645: IMP: Unhandled messages duplicate keys fix

- it has been observed that the unhandled messages are not logged correctly and turned out it is because of the Types of each parameter used as keys and keys can't be duplicate by default which caused issue for the unhandled messages, since they may have 6 parameter and all with F32 type which then becomes one parameter of type F32.

- an attribute has been defined for the config file with the format:

#--duplicate_key_on

#--duplicate_key_off

to allow or disallow the duplicate key at any point in the config file.

- Note that the save config parameters are always unique key like AlarmVolume which doesn't make sense to have duplicate entry.

  1. … 9 more files in changeset.
DEN-13645: IMP: DEN-9297: Disinfection messaging flow review

- Disinfection current code/design review

- Disinfection Screens and Navigation implementation

- There are functions and properties used but not implemented yet in vDisinfectAdjustDisinfect, and some that are implemented are not implemented with the new design.

    • -146
    • +31
    ./pages/disinfect/DisinfectStack.qml
  1. … 3 more files in changeset.
DEN-13645: IMP: DEN-13473: HD Communication/UI Issues

***** THIS IS AN EXPERIMENT *****

- It seems like the UI could be initialized early on, even before the controllers, so when the controllers are sending messages to UI to update the screen it is initialized and available to get update.

- previously when controllers were publishing the statuses UI was not even initialized and ready to receive any signal and that seems to be the reason some part of the UI (SD-Card status) was not get updated and missed signals.

_____

- Also added NoCANBus to able to go to Service when there is no HD to confirm the state change.

  1. … 2 more files in changeset.
DEN-13645 IMP: DEN-12958: UI shall generate the POST fail alarm if the SD-Card not present and stop logging

__________

- Log Export

- Added the Alarm generation on the GuiController if SD-Card is not ready, or is/becomes read-only.

- AlarmGenerator

- Added the SD-Card failure alarm - ALARM_ID_UI_POST_FAILURE_SDCARD

- Added alarm clear - ALARM_ID_NO_ALARM

- Added a flag to stop logging if the SD-Card is not ready or is readonly.

- Removed pause on log clean up.

- Improved the DeviceController SD-Card, USB status detection.

- Fixed and improved the MVC connection betweein device/Application/Gui controllers.

- A CLI switch ( -S ) added to disable the feature that is disabling the log if the SD-Card doesn't exist. So if the -S switch is used while calling the denali application, the logging will NOT be disabled, if SD-Card fails.

- In the ExportButton and USBButton(eject) removed the enable logic and has to be done outside of the component. It has been done for the Settings/export and Post-Tx/Review.

- Added title and file count for each SD-Card and USB-Drive panels.

__________

- Run.sh

- Added a check for the CloudSync and if the folder doesn't exist it won't do the rest

- Kill the denali application, in case it is still running.

- Added SD-Card type as part of its detection and if it is not ext4 format it fails the test.

- Optimized the mount point detection.

__________

- NoCANBus

- added the NoCANBus to the fault and init modes to not to navigate to those if NoCANBus is true. Otherwise it conflict with navigate to the home.

- Added a log in case the NoCANBus is enabled.

__________

- Settings

- Converted the '\n' in the settings values to actual newline character.

__________

- Thread:

- Moved the QuiController in its own thread, _Gui_Thread.

__________

- Alarms

- Added an emit to update the Alarm Dialog with the latest title/message, received from Alarms.conf In case there was an alarm before even Alarm.conf is read.

    • -6
    • +12
    ./pages/settings/SettingsExportLogs.qml
  1. … 23 more files in changeset.
DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

- Finalizing the log export.

    • -30
    • +32
    ./pages/settings/SettingsExportLogs.qml
  1. … 1 more file in changeset.
DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

- Update the eject and export button status for the export service and treatment

  1. … 2 more files in changeset.
13645 Improving NoCANBus DevConf

DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

- Changed to device folders.

    • -1
    • +1
    ./pages/settings/SettingsExportLogs.qml
DEN-13645: IMP: General Implementation

DEN-12836: Treatment Report/log export modifications

    • -21
    • +144
    ./pages/settings/SettingsExportLogs.qml
  1. … 11 more files in changeset.
DEN-13489 IMP: General Implementation

DEN-13035: Update alarm body messages

- Removed the exclamation mark icon form the Alarm Dialog.

- Removed the exclamation mark icon form the Alarm Bar.

- Made the whole AlarmDialog title bar doubleClick and diagnostics opener if the alarm is not minimizable.

- Moved the Alarm ID from title to the bottom left of the dialog with format: "ID:<id>".

- Clipped the AlarmDialon body message area, crop the text if longer than dialog area.

- Changed the "settings" property name of the VSettings to "instructions".

- Changed the date/time formats keys to simplify the code and conf file.

- Implemented the Alarm Body messages, to load from Alams/Alarms.conf, instead of the FW.

- The FW texts/documents will still being used if the Alarms.conf doesn't have a definition.

- A instruction base capability has also been added to the MVC, but UI still not supporting it.

    • -18
    • +27
    ./dialogs/NotificationDialog.qml
  1. … 13 more files in changeset.
Fixed the issue on the PatientID confirm text

  1. … 1 more file in changeset.
DEN-13489: IMP: General Implementation for any bugs and any other implementations

- Enable Dialin logging CL switch has changed to disable dialin logging, to make the dialin logging as enabled by default and removed -d from run.sh. [Sean's Request]

- Added log unhandled as error to the unhandled reader as well.

- Implemented the NoCANBus to be used when no CANBus enabled. For now it is only bypassing the create treatment button command and helps to get to the treatment parameters for the touchscreen test. [EEMN's Support]

- Fixed a bug on the Volume Adjustment on post, sending to HD to update HD alarm volume with previous, adjusted value, which was sending a request one again receiving HD response, and it was a potential crash loop.

- Added disable acknowledgment logs (req,bak) CL switch, -K : --disable-acknow-log [Sean's Request]

  1. … 13 more files in changeset.
DEN-13489: IMP: General Implementation for any bugs and any other implementations

DEN-13163: Pressing Settings and Treatment goes back to Blood Priming

- The issue was the test codes of the trending screen wasn't commented out.

DEN-13489 IMP: General Implementation

DEN-12838: UI shall store the latest alarm volume and resend that to HD on POST

DEN-11539: Service password initial setup

  1. … 10 more files in changeset.
DEN-12838: UI shall store the latest alarm volume and resend that to HD on POST

- The save ability has been added to the Settings.

- Note: the design of the settings need to change: 1- singleton class should be the controller not the mode, 2 - The column based keyvalue has to change to tree(map).

  1. … 10 more files in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- A bug found on multiple, ON,OFF toggle, slider sends minimum instead of '0' while ON.

- Slider improved.

- Added float rounding of the request message logs to 3.

    • -3
    • +4
    ./components/SliderCreateTreatment.qml
  1. … 3 more files in changeset.
DEN-13049 - UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

- There was a bug in the fix which was related to the validation and caused to fail the treatment parameters validation.

- also removed the disable if the SD is not detected correctly. (QStorageInfo is not sending DeviceReady correctly)

- enabled the float rounding to 3 in the log

- Enabled the Create Treatment button rejection notification, since it seems it is not being handled by alarms as presumed.

- Fixed the BloodPressure interval value to display OFF when set to zero.

  1. … 2 more files in changeset.
DEN-9129: Unable to proceed past "Create a Custom Treatment" when Arterial and Venous Pressure (High) Limits maximum values are initially selected

Also modified the SliderCreateTreatment.qml to show by the default the zero value not the OFF since it is only two Heparin Bolus/Dispensing which display OFF for zero value.

    • -8
    • +4
    ./components/SliderCreateTreatment.qml
  1. … 1 more file in changeset.
DEN-13049: UI Treatment Parameters Entry: Some values for treatment parameter entry on treatment parameter entry screen in create treatment do not match SRS/PRS

    • -0
    • +1
    ./components/SliderCreateTreatment.qml
  1. … 2 more files in changeset.
DEN-13199: INT: DVT Support

- During DVT support found out the BCuff vitals debug item on the top of the screen is not working, and fixed ti by adding a real time version of the same actual vital properties which gets updated every time the vitals received.

  1. … 5 more files in changeset.
DEN-13127: LTS: Development Support

- [Export Button Enabling in VM #61|https://github.com/Diality/testsuites/issues/61]

  1. … 6 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

    • -0
    • +89
    ./SDCProgressItem.qml
    • -0
    • +90
    ./USBProgressItem.qml
  1. … 30 more files in changeset.
DEN-12836: Treatment Report/log export modifications

- DEN-12944: KBM: Usual Meetings/Support/Development

- Modified the setup.sh script to work on PDT (Pacific timezone)

- Added timezone display on the screen by adding it to the VDateTime

- Removed unnecessary FileHandler header include.

- Modified the Treatment Review Export button to export the current TxReport only.

- Added the Safe Eject button for the Post-Treatment TxReport Export.

- Moved Log Folder definitions to the StorageGlobal.

- Fixed the greeting issue [https://github.com/Diality/testsuites/issues/57]

- DEN-11826: Bluetooth BloodPressure Cuff Interval - Observation

- It has completely redesigned

- DEN-9706: Blood Pressure measurement entry window closes automatically

- Auto Hide is disabled

    • -21
    • +20
    ./pages/treatment/TreatmentStack.qml
  1. … 21 more files in changeset.
DEN-12558: HD does not transition to service mode

- implemented the service mode confirmation

- decreased the dialog animation duration

- renamed the service dates MVC to differentiate with service mode.

    • -15
    • +27
    ./pages/settings/SettingsStack.qml
  1. … 30 more files in changeset.
DEN-12541: Get the SD-Card size from the system

  1. … 2 more files in changeset.
DEN-12804: DOX: General review and support

- Removed the compatibility version from the version format

  1. … 3 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

  1. … 22 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 151 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 12 more files in changeset.
DEN-12514: LTS: Support/Development

========== DevTest

- Fixed the reported bugs from the LTTS team:

+ https://github.com/Diality/testsuites/issues/38: Rejection messages are not displayed for heparin under main-treatment screen 38

+ https://github.com/Diality/testsuites/issues/37: "Self Test Complete!" message is not displayed when "DRY_SELF_TESTS_COMPLETE_STATE" is simulated #37

+ https://github.com/Diality/testsuites/issues/36: Indicators are not displayed for 'Wet Self tests" when "HD_PRIME_WET_SELF_TESTS_STATE" is simulated #36

========== CloudSync

- Changed the order of the device registration request and the device info send, in the order above.

========== Internal

- removed the emoji and language globe buttons from the second keyboard layout.

  1. … 9 more files in changeset.