DEN-6372: Added treatment stop states enum.

Corrected.

Corrected.

Done.

Done.

Done.

Done.

Done.

Done.

If not yet in fill mode, will dgCmdResp persist to next time?

If not yet in fill mode, will dgCmdResp persist to next time?

Is this response persistent? Or does this function consume the response? Even if persistent, I suspect a new response will replace previous response and this makes me nervous. I think it would be s...

Is this response persistent? Or does this function consume the response? Even if persistent, I suspect a new response will replace previous response and this makes me nervous. I think it would be safer to have a separate flag for each possible response so there is no race condition.

This flag does not appear to be a "result". Rename to primeCompleted.

This flag does not appear to be a "result". Rename to primeCompleted.

This is now included in the header file so we can remove here.

This is now included in the header file so we can remove here.

This function is too big. Create static functions to handle these switch cases.

This function is too big. Create static functions to handle these switch cases.

Rate is mL/min, not mL.

Rate is mL/min, not mL.

DEN-6372: Removed some redundant group end markers.

Added blood prime sub-mode states enum. Added new blood prime volume check alarm. Added new blood prime status message.

DEN-6372: Added user messaging for treatment end sub-mode.

Fixed merge conflicts.

DEN-6372: Added new Treatment End sub-mode module.

Merge branch 'Sprint37' into develop

DEN-6402: Added new alarm for no cartridge loaded

When we run blood pump in forward direction, we should see negative on arterial and positive on venous. This limit here is to stop blood pump even if the time to run the blood has not been reached.

When we run blood pump in forward direction, we should see negative on arterial and positive on venous.
This limit here is to stop blood pump even if the time to run the blood has not been reached.

Done. Also moved the function to FPGA module.

Done. Also moved the function to FPGA module.

Done.

Done.

Done.

Done.

Reworded.

Reworded.

Renamed.

Renamed.

If the sensor is bad, we will go to fault. In this case, the calibration does not matter anymore.

If the sensor is bad, we will go to fault. In this case, the calibration does not matter anymore.

Agree. All multi-lines state have been moved to a static function.

Agree. All multi-lines state have been moved to a static function.

Done.

Done.

Done. It is a tolerance on pressure difference after we return back to normal state from pressure self-tests.

Done. It is a tolerance on pressure difference after we return back to normal state from pressure self-tests.

Done. This is not a self-test for blood pump occlusion sensor (OB). This is for PBa and PBo pressure sensors.

Done. This is not a self-test for blood pump occlusion sensor (OB). This is for PBa and PBo pressure sensors.