dd-firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

merged LDT-4410 Into develop

merged latest staging

    • -41
    • +50
    /firmware/App/Controllers/MixingControl.c
    • -1
    • +5
    /firmware/App/Controllers/MixingControl.h
    • -74
    • +99
    /firmware/App/Monitors/Conductivity.c
0.0F(float type).

0.0F(float type).

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Merge branch 'LDT-4410-dialysate-composition-mixing-2' into develop

LDT-4410 commented temperature alarm for develop branch working

Stepper driver firmware implementation

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

merged LDT-4410 to develop

    • -59
    • +159
    /firmware/App/Controllers/DryBiCart.c
    • -10
    • +8
    /firmware/App/Controllers/MixingControl.c
Delete it.

Delete it.

LDT-4410 code review comments addressed

    • -36
    • +32
    /firmware/App/Controllers/DryBiCart.c
    • -7
    • +8
    /firmware/App/Controllers/MixingControl.c
Remove blank spaces after =.

Remove blank spaces after =.

To be deleted this line?

To be deleted this line?

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as w...

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as well.

I guess we can remove the "balancingChamberFillInProgress" check here.

I guess we can remove the "balancingChamberFillInProgress" check here.

Driver update code part 1

    • -37
    • +306
    /firmware/App/Services/FpgaDD.c
Remove the TODO comment.

Remove the TODO comment.

align the comment.

align the comment.

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pre...

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pressure etc.,

decays to threshold level (4 PSI).

decays to threshold level (4 PSI).

remove all commented code.

remove all commented code.

Update comment. remove TODO if no more applicable.

Update comment. remove TODO if no more applicable.

FillWaterEndState?

FillWaterEndState?

Delete it.

Delete it.

Supply state (instead of drain). Updates Inputs and outputs across the file.

Supply state (instead of drain). Updates Inputs and outputs across the file.

When the BC fill completed and allow opening D65 for Chamber F fill?

When the BC fill completed and allow opening D65 for Chamber F fill?

Can we rename this IsDryBicartChamberFillWindowOpen? ( removing BalChamber)

Can we rename this IsDryBicartChamberFillWindowOpen? ( removing BalChamber)