Driver update code part 1

  1. … 2 more files in changeset.
Remove the TODO comment.

Remove the TODO comment.

align the comment.

align the comment.

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pre...

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pressure etc.,

decays to threshold level (4 PSI).

decays to threshold level (4 PSI).

remove all commented code.

remove all commented code.

Update comment. remove TODO if no more applicable.

Update comment. remove TODO if no more applicable.

FillWaterEndState?

FillWaterEndState?

Delete it.

Delete it.

Supply state (instead of drain). Updates Inputs and outputs across the file.

Supply state (instead of drain). Updates Inputs and outputs across the file.

When the BC fill completed and allow opening D65 for Chamber F fill?

When the BC fill completed and allow opening D65 for Chamber F fill?

Can we rename this IsDryBicartChamberFillWindowOpen? ( removing BalChamber)

Can we rename this IsDryBicartChamberFillWindowOpen? ( removing BalChamber)

50ms?

50ms?

Add comment to align with the rest of the implementation.

Add comment to align with the rest of the implementation.

Add outputs as well.

Add outputs as well.

I guess we should have some minimum speed as well as like max speed. Please work with Abhijit on worst case min speed.

I guess we should have some minimum speed as well as like max speed. Please work with Abhijit on worst case min speed.

LEAHI-DD-FIRMWARE-LDT-4500_(DD) Alarm 75 and pressure build up during Chamber H fill
LEAHI-DD-FIRMWARE-LDT-4500_(DD) Alarm 75 and pressure build up during Chamber H fill
LDT-4410 made changes to state machine and cpontrol for conductivity spikes

  1. … 2 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-5186_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 7/1:
LEAHI-DD-FIRMWARE-LDT-5186_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 7/1:
LDT-5186 drain implementation with cencelation support

code re-organisation

  1. … 2 more files in changeset.
LDT-4614 fixing compilation issue after merge

Initial Changes

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch LDT-4614-dialysate-composition-mixing-algo-integration-with-td-proxy into develop

Merge branch LDT-4614-dialysate-composition-mixing-algo-integration-with-td-proxy into staging

LEAHI-DD-FIRMWARE-LDT-4611_HDF - FW Implementation - 5/4:
LEAHI-DD-FIRMWARE-LDT-4611_HDF - FW Implementation - 5/4:
LEAHI-DD-FIRMWARE-LDT-5157_System Data and Logs (DD) - Conductivity Sensor Calibration (1/3)