Use the get Acid mix control interval?

Use the get Acid mix control interval?

Updates the comments for all new variables.

Updates the comments for all new variables.

Remove this flag.

Remove this flag.

Remove this code.

Remove this code.

There should be a \b before the type of details (Inputs and Outputs here) for documentation please update for all

There should be a \b before the type of details (Inputs and Outputs here) for documentation please update for all

Remove blank line

Remove blank line

Remove this condition to apply the speed for BC switch only as well.

Remove this condition to apply the speed for BC switch only as well.

Remove blank line There should be a \b before the type of details (Inputs and Outputs here) for documentation

Remove blank line
There should be a \b before the type of details (Inputs and Outputs here) for documentation

There should be a \b before the type of details (Inputs and Outputs here) for documentation Inputs: None?

There should be a \b before the type of details (Inputs and Outputs here) for documentation
Inputs: None?

Check the else portion and do the required changes.

Check the else portion and do the required changes.

Name something similar to 'BCSwitchingBasedOnClosePeriodCounter'?

Name something similar to 'BCSwitchingBasedOnClosePeriodCounter'?

I guess, all these checks should be moved to FillEndState.

I guess, all these checks should be moved to FillEndState.

after one state of BC switching completed. update the next comment as well.

after one state of BC switching completed. update the next comment as well.

LDT-3963 dialysate flow rate algorithm changes and testing

  1. … 8 more files in changeset.
Should we add a TODO to remove this one as well?

Should we add a TODO to remove this one as well?

If we are removing this call, then where are we calling the function testDDFloaterLevelStateOverride?

If we are removing this call, then where are we calling the function testDDFloaterLevelStateOverride?

Please update the function name : testDDFloaterLevelStateOverride

Please update the function name : testDDFloaterLevelStateOverride

LDT-4183 drybicart firmware implementation

Merge branch LDT-4183-dialysate-composition-dry-bicar into develop

  1. … 1 more file in changeset.
LDT-4183 addressed self review comments plus dry bicarb implementation

Why are we removing this TODO comment? It looks like we should keep it.

Why are we removing this TODO comment? It looks like we should keep it.

Do we need TODO comments to restore these alarms later?

Do we need TODO comments to restore these alarms later?

What is the point of this message?

What is the point of this message?

Looks like we need a TODO comment to restore these checks later.

Looks like we need a TODO comment to restore these checks later.

LEAHI-DD-FIRMWARE-LDT-4183_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 2/1: