Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15265 addressed a code review comment

Should be &result.

Should be &result.

DEN-15265, DEN-15236 changed the chemical disinfect monitor to check the dialysate cap not to be closed in during the disinfect since the PAA acid is connected to it.

Merged DEN-15265

DEN-15265, DEN-15246 changed the heat disinfect's primary target temperature to 60 C prior to filling with water for the actual heat disinfct. Updated the code to properly exit the bad fill state of the gen idle mode once the bad sequence is finished

DG-DEN-15265_Heat Disinfect Actuator Requirements Contain Outdated Information For Heater Target
DG-DEN-15265_Heat Disinfect Actuator Requirements Contain Outdated Information For Heater Target
This looks like 104.

This looks like 104.

Probably not needed, but for good measure can we do this in transition to GenIdle mode too?

Probably not needed, but for good measure can we do this in transition to GenIdle mode too?

Add blank line before return statement.

Add blank line before return statement.

This function could, in theory, be used to assign a CD2 cal table to CPo (for example).

This function could, in theory, be used to assign a CD2 cal table to CPo (for example).

DG-DEN-14902_Bootloader For DG
DG-DEN-14902_Bootloader For DG
Merge branch 'DEN-15246-active-reservoir-empty-post-concentrate-alarm-recovery' into develop

  1. … 2 more files in changeset.
DEN-15246 added more code to support different conductivity sensors calibration tables

  1. … 4 more files in changeset.
Update header.

Update header.

DG-DEN-15246_Active Reservoir Empty Post Concentrate Alarm Recovery
DG-DEN-15246_Active Reservoir Empty Post Concentrate Alarm Recovery
DEN-15246 changed the bad fill end of state

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15231

  1. … 4 more files in changeset.
DG-DEN-15231_Alarm 89 Uvo ON NO Flow Triggers During Treatment
DG-DEN-15231_Alarm 89 Uvo ON NO Flow Triggers During Treatment
DEN-15231 changed the target flow in the mode drain to 300 mL/min and the drain pump's target RPM to 2200

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

I said fix throughout. Need fix in function below as well.

I said fix throughout. Need fix in function below as well.