Are these permanent changes?

Are these permanent changes?

Comment doesn't match test flag.

Comment doesn't match test flag.

DG-DEN-17234_Alarm 68 Fix IN Expedite Pretreatment
DG-DEN-17234_Alarm 68 Fix IN Expedite Pretreatment
DG-DEN-17220_Release 20 P1B Staging
DG-DEN-17220_Release 20 P1B Staging
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Updated the trigger condition. This is a calibration and is checked in POST. The titles are POST.

Updated the trigger condition. This is a calibration and is checked in POST. The titles are POST.

The Trigger condition is little confusing. could you please reword it? Also In title, do you want to add " fault", if it is DG fault property?

The Trigger condition is little confusing. could you please reword it? Also In title, do you want to add " fault", if it is DG fault property?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed the typecast.

Removed the typecast.

Are these castings of (POLYNOMIAL_CAL_PAYLOAD_T*) necessary? Seems like &concPump->concentratePumps[ i ] is already that type.

Are these castings of (POLYNOMIAL_CAL_PAYLOAD_T*) necessary? Seems like &concPump->concentratePumps[ i ] is already that type.

DG-DEN-17135_Concentrate Pump Calibration Staging
DG-DEN-17135_Concentrate Pump Calibration Staging
DG-DEN-17093_Blood Leak Zeroing IN Treatment P1B Staging
DG-DEN-17093_Blood Leak Zeroing IN Treatment P1B Staging
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

I think so. This is to catch all the time that the concentrate pumps are running. If the concentrate pumps are not running, then nothing is accumulated.

I think so. This is to catch all the time that the concentrate pumps are running. If the concentrate pumps are not running, then nothing is accumulated.

Do we what to do this in ALL fill states?

Do we what to do this in ALL fill states?

DG-DENBUG-276_Alarm 282 Bicarb Usage Alarm Did Not Trigger Before The Bicarb Ran Out
DG-DENBUG-276_Alarm 282 Bicarb Usage Alarm Did Not Trigger Before The Bicarb Ran Out