DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Corrected comment alignment

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Undated 5second timeout to 10 seconds

DEN-15697 changed the code to correctly set the drain timeout in active cool mode

DEN-15619 removed the variable to not trigger a temperature not reached to target alarm in chemical disinfect

DEN-15660, DEN-15523, DEN-15528 changed the acid deprime state in chemical disinfect. Updated the chemical disinfect prime state

DEN-15660, DEN-15727 removed the chemical disinfect flush complete date set from its complete state since it should only be updated if the sample passed

DEN-15660, DEN-15528, DEN-15523 updated the chemical disinfect's moving average. Added a new state to deprime the acid line in chemical disinfect and then park the acid pumpf for the rest of the chemical disinfect

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15701: handle stop DG command from HD while in standby water sampling states.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15647 addressed the code review comments

DEN-15610 fixed DG flush to write basic flush complete timestamp in the usage record

Merged DEN-15647

  1. … 1 more file in changeset.
DEN-15647 updated the heat disinfect active cool mode to deenergize the actuators prior to transitioning to cancel water path state

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15571, DEN-15398 changed the fill mode full speed prime time from 8 seconds to 7 seconds

DEN-15571 addressed the code review comments

DEN-15571 updated the chemical disinfect drain status function

DEN-15571 made changes in the chemical disinfect mode to not trigger the max temeprature and conductivity out of range until the end of the chemical disinfect mode

Merged staging

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
DEN-15571 changed the chemical disinfect temperature timeout from 1 minute to 5 minutes

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 4 more files in changeset.
DEN-15367 removed the payload from the message that informs HD that DG is ready to dispense more fluid in DG. DEN-15571 fixed the chemical disinfect state machine to transition to cancel water path once the target temperature was not reached within the specified time

  1. … 1 more file in changeset.
DEN-15367 addressed the code review comments

  1. … 5 more files in changeset.
DEN-15522 fixed the heat disinfect active cool fill timeout not triggering. DEN-14805 fixed chemical disinfect reservoir leak alarm

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 4 more files in changeset.
DEN-15367 changed the RO permeate sample cancel water path. DEN-14004 changed the request message from UI to 0x89. DEN-15509 changed the reservoir threshold from 1700 to 1850 mL. DEN-15192 changed the concentrate pump low speed threshold.

  1. … 2 more files in changeset.
-DEN-15367 fixed the transition to cancel water path in RO permeate sample mode -DEN-14004 changed the request message ID to 0x89 from UI to send the DG service record -DEN-15192 changed the concentrate pumps low speed check threshold -DEN-15509 changed the reservoir full threshold from 1700 mL to 1850 mL.

  1. … 3 more files in changeset.
DEN-14004 added another message to receive and send DG service record upon request from HD. The format of the payload is different from the format that is sent to UI

  1. … 3 more files in changeset.