Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-335: air trap fill refinements.

  1. … 4 more files in changeset.
DENBUG-335: First cut at new air trap fill approach with closing of VBV and slowing of BP.

  1. … 10 more files in changeset.
HD-DENBUG-335_Air Trap Fill Design Improvem
HD-DENBUG-335_Air Trap Fill Design Improvem
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 118 more files in changeset.
Merged release/p1b/staging

  1. … 9 more files in changeset.
DENBUG-334 added the changes from staging

  1. … 9 more files in changeset.
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle P1B Staging
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle P1B Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 118 more files in changeset.
DENBUG-334 updated the code to initialize the variables in between the treatments without a need to power cycle the unit

  1. … 9 more files in changeset.
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
Merged release/p1b/staging

  1. … 4 more files in changeset.
DENBUG-328 added code from staging

  1. … 4 more files in changeset.
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-328 updated the air trap event and the VectorCAST test cases

DENBUG-328 changed the get raw air trap function to get air trap function for alarm 215. Added air trap state change events

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.