BloodFlow.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Fixes per UT and IT. Updates to UT and IT.

  1. … 7 more files in changeset.
Fixes for UF and treatment time changes. Updated pump factors.

  1. … 9 more files in changeset.
Changed breadboard build switch to be more biased toward the new HD board. Added UF pause resume functions and message handling. Updated internal ADC voltage and conversion factors for new HD board. Commented out RTC POST until RTC on new HD board is configured.

  1. … 12 more files in changeset.
Added HD specific common header. Streamlined includes for Common.h and HDCommon.h.

  1. … 55 more files in changeset.
CAN Test #1.

  1. … 8 more files in changeset.
Updated UT for pump modules to include open loop mode.

  1. … 5 more files in changeset.
Added dialysis sub-mode. Misc fixes.

  1. … 25 more files in changeset.
Added DPo implementation. Added some messagine for DPo. Added FIRFilters. Updated HALCoGen pinouts.

  1. … 38 more files in changeset.
Added function to pump modules to set the PWM. Minor fixes to CAN ACK feature. Started updated UT/IT for CAN ACK feature.

  1. … 5 more files in changeset.
Added new PresOccl, Dialysis, and TreatmentStop modules. Updated test scripts for system communication modules.

  1. … 16 more files in changeset.
Tuned pump control. Fixed PIController anti-windup. Added some temporary debug code. Added feature to CAN for messages that require an ACK.

  1. … 15 more files in changeset.
Merge branch 'master' into DEN-1312-hd-dialysate-inlet-pump-drive-flow-sensor-control

  1. … 28 more files in changeset.
Updated PIControllers to behave in classic manner. Added more thread protection to CAN comm buffers. +review HD-DEN-1312-1

  1. … 8 more files in changeset.
Updated BloodFlow and DialInFlow modules per UT. Added new UT for DialInFlow.

  1. … 3 more files in changeset.
Removed DialOutFlow module - Leo added to separate branch. Fixed alarm lamp POST.

  1. … 18 more files in changeset.
Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 64 more files in changeset.
Added PIControllers module. Used new PIControllers module in BloodFlow and DialInFlow modules. Moved Watchdog POST to first POST test. Prevent watchdog expired fault during watchdog POST test.

  1. … 5 more files in changeset.
Updated CAN interrupts config. Added a watchdog expired alarm. Added PWM duty cycle to blood flow broadcast.

  1. … 18 more files in changeset.
Added some more CAN debug code. Minor updates to blood and dialysate pump modules.

  1. … 5 more files in changeset.
Added SPI driver for GPIO. Set appropriate motor dir and stop pins for dialysate pump modules. Added calls to init dialysate pump modules. Added exec calls from tasks for dialysate pump modules.

  1. … 10 more files in changeset.
Updated flow filter in BloodFlow per discussion with Blaine. Added ADC channels for dialysate pumps. Added some debug code for CAN tranmit issue. Added GPIO for dialysate pumps to HALCoGen.

  1. … 19 more files in changeset.
Added bloodflow overrides for not yet implemented sensors. Condensed published alarms status message.

  1. … 6 more files in changeset.
Code review fix. Adding rounding to BloodFlow module. +review HD-DEN-759-1

  1. … 67 more files in changeset.
Fixed white space between brackets throughout. Added some debug code to help troubleshoot comm issue. Updated power off messaging. +review HD-DEN-759-1

  1. … 84 more files in changeset.
Added place holders for 2 sensors not yet ready for BloodFlow module. Refactored IT for CPLD and SafetyShutdown modules to move from target run to MinGW run tests. Updated UT for SystemComm and SystemCommMessages modules. +review HD-DEN-759-1

  1. … 3 more files in changeset.
Added UT for BloodFlow module. Updated other UT. Added IT for AlarmMgmt and BloodFlow modules. Updated other IT. +review HD-DEN-759-1

  1. … 6 more files in changeset.
Started new UT for BloodFlow module. Finished UT for AlarmMgmt module. Fixed issues discovered during UT. +review HD-DEN-759-1

  1. … 5 more files in changeset.
Code review fixes. +review HD-DEN-759-1

Addressed code review feedback. Fixed an intermittent CAN bug. Updated UT for Buttons, CPLD, and Interrupts modules. +review HD-DEN-759-1

  1. … 11 more files in changeset.
Addressed some code review feedback. Added some checks to BloodFlow monitor function per requirements. +review HD-DEN-759-1

  1. … 3 more files in changeset.