DEN-14007: Tentative fix for DEN-13688 for pump homing consistency - set all pumps to same speed while homing.

HD-DEN-14007_2 Change Request DE Prime Dialysate Lines
HD-DEN-14007_2 Change Request DE Prime Dialysate Lines
DEN-14497, DEN-14326 Added a function to regularly schedule the informative blood leak embedded mode commands. Changed the FALSE to ACk not required in some of the message ack functions

  1. … 3 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

DEN-14344: addressed code review comments

  1. … 1 more file in changeset.
For a release build, looks like we would do nothing and get stuck in this state.

For a release build, looks like we would do nothing and get stuck in this state.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 1 more file in changeset.
HD-DEN-14497_DG HD Dev HD DG Dvt Update Part 7
HD-DEN-14497_DG HD Dev HD DG Dvt Update Part 7
DEN-14344: removed venous high limit increase while air pump is operating. Increased air pump operation timeout to 6s to handle worst case bp ramping.

DEN-14344: changed air pump upper level persistence from 2s to 100ms.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14007 merged staging and resolved the merge conflicts

  1. … 1 more file in changeset.
DEN-14007, DEN-14099 Merge & fix conflicts.

  1. … 1 more file in changeset.
Fixed. Thanks!

Fixed. Thanks!

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14344: code review fixes.

DEN-14344: code review fixes.