Added safety shutdown override command support. Zero treatment duration param in response to UF adjustment message if time adjustment should be disabled.

  1. … 6 more files in changeset.
Perparing for a merge with DEN-1906

  1. … 43 more files in changeset.
Changed breadboard build switch to be more biased toward the new HD board. Added UF pause resume functions and message handling. Updated internal ADC voltage and conversion factors for new HD board. Commented out RTC POST until RTC on new HD board is configured.

  1. … 12 more files in changeset.
Added HD specific common header. Streamlined includes for Common.h and HDCommon.h.

  1. … 51 more files in changeset.
Put all debug UART code inside a debug build switch. Misc. other minor changes.

  1. … 18 more files in changeset.
Added DPo implementation. Added some messagine for DPo. Added FIRFilters. Updated HALCoGen pinouts.

  1. … 38 more files in changeset.
Merge branch 'master' into DEN-1312-hd-dialysate-inlet-pump-drive-flow-sensor-control

  1. … 26 more files in changeset.
Minor improvements to CAN communications. +review HD-DEN-1312-1

  1. … 3 more files in changeset.
Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 57 more files in changeset.
Updated flow filter in BloodFlow per discussion with Blaine. Added ADC channels for dialysate pumps. Added some debug code for CAN tranmit issue. Added GPIO for dialysate pumps to HALCoGen.

  1. … 18 more files in changeset.
Increased delay between CAN bus recovery attempts. Added a little handling for UART frame/overrun errors. Updated UT. +review HD-DEN-759-1

  1. … 4 more files in changeset.
Fixed white space between brackets throughout. Added some debug code to help troubleshoot comm issue. Updated power off messaging. +review HD-DEN-759-1

  1. … 83 more files in changeset.
Started new UT for BloodFlow module. Finished UT for AlarmMgmt module. Fixed issues discovered during UT. +review HD-DEN-759-1

  1. … 5 more files in changeset.
Addressed code review feedback. Fixed an intermittent CAN bug. Updated UT for Buttons, CPLD, and Interrupts modules. +review HD-DEN-759-1

  1. … 9 more files in changeset.
Initial commit of the Real Time Clock code

  1. … 11 more files in changeset.
Added blood pump control per flow. Added timeout to power off request. Added more interrupt handling. Added parity checking for CAN RAM. +review HD-DEN_759-1

  1. … 12 more files in changeset.
Misc fixes to BloodFlow and AlarmMgmt modules. Scaled BloodFlow analogs. Changed cargo to payload throughout code. Updates to existing UT. +review HD-DEN-759-1

  1. … 14 more files in changeset.
Updated UT for AlarmLamp, Comm, CPLD, and SafetyShutdown modules. +review HD-DEN-759-1

Addressed code review feedback. +review HD-DEN-759-1

  1. … 11 more files in changeset.
Replaced TODOs for triggering s/w faults with actual triggers of s/w faults. +review HD-DEN-759-1

  1. … 16 more files in changeset.
Added InternalADC module. Configured ADC peripheral. +review HD-DEN-759-1

  1. … 13 more files in changeset.
Added BloodFlow module. Added ADC and PWM peripherals. Fixed #includes.

  1. … 32 more files in changeset.
Transferred the hd_sprint5 firmware to the firmware folder

  1. … 392 more files in changeset.
Transferred the hd_sprint4 firmware to the firmware folder

  1. … 392 more files in changeset.
Transferred the sprint3 firmware to the firmware folder

  1. … 376 more files in changeset.
Transferred the master branch firmware to the firmware folder. Added the VectorCAST folder to the repository

  1. … 392 more files in changeset.