td-firmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Added doxygen

    • -41
    • +40
    /firmware/App/Controllers/Valves.c
Can't return in middle of function code.

Can't return in middle of function code.

Remove else and reject command in it and re-test.

Remove else and reject command in it and re-test.

Added doxygen

Added doxygen

    • -112
    • +112
    /firmware/App/Drivers/PinchValve.c
New changes

    • -68
    • +83
    /firmware/App/Drivers/PinchValve.c
    • -16
    • +18
    /firmware/App/Drivers/PinchValve.h
LDT-4049:Addressed review comments

Changes

    • -40
    • +84
    /firmware/App/Drivers/PinchValve.c
Changes

    • -29
    • +35
    /firmware/App/Drivers/PinchValve.c
Merge branch 'LDT-4149-pinch-valves-fw-implementation' of bitbucket.org:diality-cloud/td-firmware into LDT-4149-pinch-valves-fw-implementation

New Add ons

    • -140
    • +266
    /firmware/App/Drivers/PinchValve.c
LDT:4063- addressed review comments and updated the shutdown flag prompt

LDT-4029:Addressed review comments

    • -14
    • +27
    /firmware/App/Controllers/SyringePump.c
    • -0
    • +1
    /firmware/App/Controllers/SyringePump.h
LDT-4149: Initialized ABC valve positions array.

break this line

break this line

recircPublishTimerCtr is an output

recircPublishTimerCtr is an output

Add @message MSG_ID_TD_RECIRCULATE_RESPONSE

Add @message MSG_ID_TD_RECIRCULATE_RESPONSE

break this line

break this line

Initialization missing for the following variables in the init function. startRecircRequested reconnectRequested backToTreatmentRequested resumeRecircRequested

Initialization missing for the following variables in the init function.

startRecircRequested
reconnectRequested
backToTreatmentRequested
resumeRecircRequested

All signal flags are initialized in resetSignalFlags() which we are calling below

All signal flags are initialized in resetSignalFlags() which we are calling below

LDT-4149: more changes.

please update the alarm id reservation sheet.

please update the alarm id reservation sheet.

Why is MSG_ID_DD_SUBSTITUTION_PUMP_DATA commented out? Line 214 and 215 are using tabs instead of whitespace characters because of which the alignment is off.

Why is MSG_ID_DD_SUBSTITUTION_PUMP_DATA commented out?
Line 214 and 215 are using tabs instead of whitespace characters because of which the alignment is off.

break this line : one variable per line will be easy to read.

break this line : one variable per line will be easy to read.

StateTxRecirc should be above StateTxRinseback

StateTxRecirc should be above StateTxRinseback

Add @Alarm ALARM_ID_TD_SALINE_BAG_VOLUME_LOW_ALARM

Add @Alarm ALARM_ID_TD_SALINE_BAG_VOLUME_LOW_ALARM

break this line

break this line

break this line

break this line

Add rinsebackPublishTimerCtr to outputs

Add rinsebackPublishTimerCtr to outputs