Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LEAH-838: Air trap control looking to maintain level at upper level.

  1. … 1 more file in changeset.
LEAH-838: Fix blood pump stop command. Fix raw level sensor readings. Added air trap level control test code build switch (temporary code for special test build).

  1. … 5 more files in changeset.
LEAH-838: Finished implementing design change to air trap level control using 2 3-way valves and an air pump.

  1. … 2 more files in changeset.
LEAH-838: Deleted 2-way valve driver. Added 3-way valve driver per air trap level control design change. Changed air pump control from simple on/off to variable power level control.

  1. … 12 more files in changeset.
TEST: refactor component names.

  1. … 18 more files in changeset.
TEST: refactor of component names.

  1. … 18 more files in changeset.
LEAH1970: Refactored component codes to new style of codes for consistency.

  1. … 20 more files in changeset.
TEST: refactor component codes to new codes for consistency.

  1. … 20 more files in changeset.
LEAH-1970: Increased BP ramp rate. Misc cosmetic fixes.

TEST: misc cosmetic fixes.

TEST: Changed BP ramp rate to 1000 RPM per second.

TEST: changes during alpha prototype integration.

  1. … 14 more files in changeset.
TEST: compensate commanded BP speed so that it will match measured speed.

  1. … 3 more files in changeset.
TEST: Fixed blood speed feedback issue - coming back as a period instead of RPM. Fixed pressure conversion.

  1. … 5 more files in changeset.
LEAH-39: Fixed include issue.

LEAH-51: Addressed code review comments.

LEAH-51: Further developed BloodFlow unit. Addressed compiler errors/warnings.

  1. … 4 more files in changeset.
LEAH-51: Addressed code review comments. Added 2 valve fault alarms.

  1. … 4 more files in changeset.
LEAH-51: Addressed code review comments.

  1. … 6 more files in changeset.
LEAH-51: Addressed code review comments.

  1. … 1 more file in changeset.
LEAH-51: Addressed code review comments.

LEAH-51: finish valve state machine.

Added air trap, level sensors, valves.

  1. … 25 more files in changeset.
LEAH-47: Addressed code review feedback.

  1. … 2 more files in changeset.
LEAH-47: Misc fixes.

  1. … 4 more files in changeset.
LEAH-47: Added driver and controller for air pump. Added buttons monitor. Added switches monitor. Used new override message helper functions.

  1. … 22 more files in changeset.
Addressed code review comments.

  1. … 2 more files in changeset.
Added read and error count check for pressure sensor driver. Touched all units for code review purposes.

  1. … 46 more files in changeset.
split out alarm audio into separate unit.

Updated function headers.

  1. … 45 more files in changeset.