pretreatment

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-13845: CRV: Merge issue of DEN-12716-UI_BN_S70 into staging

- Due to some enum changes, UI couldn't build

- An idea implemented to detect the enum changes by compiler. The implementation may be able to improve.

  1. … 12 more files in changeset.
DEN-13134: IMP: General Implementation for any bugs and any other implementations

DEN-12836: Treatment Report/log export modifications

- Improved the SDItem and renamed to SDCProgressItem.

- Added another separate progress circle component for the USB Driver [USBProgressItem]

- Adding Space check for USB as well to get the USBProgressCircle update.

- Optimized ExportButton and added two size for Settings screen(normal) and Treatment Review screen(small)

- Improved USB button comonent

- Added a global byte size converter in Variable.qml

- Improved MainMenu component

- Added a notification for the TreatmentFlowBase.qml main component.

- Used the notification of the TreatmentFlowBase.qml for the Treatment Review Screen Log export to display feedback to the user.

- Added an individual screen for the log export.

- Removed the Export Log and USB eject from the main settings option screen.

- Added an option for the Export log on the settings home screen which is currently enabled for non-service users too.

- Fixed the DGVersionResponse, log message.

- Put a comment in the FileHandler.cpp about the Qt API QStorageInfo::mountedVolume() which seems so slow on detecting mounted volumes when a USB derive is just plugged in (It could be system issue, since it has also been observed from shell console.)

- Added error handling to the Treatment Log/Report export

DEN-12726: Denali application crashing

- Modified the Logger Maxusage limit to keep 50% of the data/event log and 15% of the service log and keep 35% free, after the log clean up triggered by 15% free storage detected.

DEN-13051: Encounter alarm 6 and SD card detection problem after updating UI version

DEN-13156: Settings screen menu should disappear on service mode

- Improved the MainStack.qml navigation rules.

- Improved the Patient ID keyboard handling, to hide the keyboard, when not on the screen.

- Improved the SettingsStack.qml navigation rules.

DEN-13128: KBM: Cloud Sync Integration

- Saving the CS log in UTC timezone.

DEN-13127: LTS: Development Support

- Fixed the issue found on the End-Treatment screen Confirm button which didn't respect the last instruction page enable only rule.

  1. … 44 more files in changeset.
DEN-12836: Treatment Report/log export modifications

- DEN-12944: KBM: Usual Meetings/Support/Development

- Modified the setup.sh script to work on PDT (Pacific timezone)

- Added timezone display on the screen by adding it to the VDateTime

- Removed unnecessary FileHandler header include.

- Modified the Treatment Review Export button to export the current TxReport only.

- Added the Safe Eject button for the Post-Treatment TxReport Export.

- Moved Log Folder definitions to the StorageGlobal.

- Fixed the greeting issue [https://github.com/Diality/testsuites/issues/57]

- DEN-11826: Bluetooth BloodPressure Cuff Interval - Observation

- It has completely redesigned

- DEN-9706: Blood Pressure measurement entry window closes automatically

- Auto Hide is disabled

    • -4
    • +11
    ./connection/PreTreatmentConnectionStack.qml
  1. … 26 more files in changeset.
wrapping up multiple sprint codes

- LOG: Added a CLI option to disable the Check-In logs by adding -k, --disable-check-in-log at the CLI. This was interrupting the log tests. The build macro has been removed.

- POST: During the coverage find out that the getter functions have not even been called in the ApplicationPost.

- CloudSync: Added the Treatment Code(TxCode), received from the cloud after the Treatment Report stored by Cloud System. It is a message communication between UI<=(2008)=CS.

- CloudSync: Two error added for the 2008: TxCode

- CloudSync: Due to some issues on Cloud System regarding the credential expiration, It has been asked to add a temporary workaround to send the credentials (1004) before message 1006:Device State, 1007:Treatment Report.

- DEN-12727: Treatment Parameter Screen: Added more space between the parameters components specially the Temperature slider and the option on top.

- DEN-12829: Enabled the back button on the Start Treatment Screen

- DEN-12761: Enabled an empty Patient ID.

It has not yet been implemented on the CloudSync App, so if sent empty CloudSystem will later reject to store the TxReport.

When later implemented an unassigned random Patient ID will be created for that TxReport to be later defined and assigned.

- DEN-12837, DEN-12840: The Disinfection Scheduling Entry and the Confirm button have been removed from the "Dialysate Generator Settings" screen.

- DEN-12837, DEN-12840: The "Dialysate Generator Settings" changed to "DG Cleaning" for the menu option and the screen title.

- Settings: The Clear Alarm Condition button has been removed from the settings screen for the release.

- Settings: The options which has no implementation for the phase 1 have been removed from the Settings screen Service mode options. [Language, Calibration, SW Update, Factory Reset]

- DEN-12841: Reworded the password entries in Service screen and for WiFi Join Screen.

- TxCode, TxReport: removed the commented out Treatment Report Parameters from the code.

- LOG: added one character space int he DEBUG log for the Not Enough Data in Types.

- Settings: The Service Dates default values changed from 0 to "" since they are strings.

- Settings: Added some clarification for the Home screen greetings

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 29 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 159 more files in changeset.
DEN-12512: INT: Support/Meetings/Planning

DEN-09712 Date format between system date/time and BP/HP "Last Read" is not consistent

DEN-09929 Treatment Log shows "?C" instead of "°C"

PRS-35: Regarding Sean's point about the create treatment parameters, made the parameters dynamic by the Configurations/DataList.cfg in ui.config repository.

  1. … 14 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 74 more files in changeset.
DEN-12180: Internal general support

- investigate on the Begin Prime back to Cartridge installation

- Added objectName for easier debug.

- added the stack change for the Cartridge Installation in case FW wants to navigate back from Begin Prime to the beginning of Cartridge Installation.

  1. … 3 more files in changeset.
DEN-10974: Syringe pump installation screen allows user to proceed before syringe installed

- Added the response message 0xAD00(173), for the request message of 0x6200(98):Pre-treatmetn disposables confirm.

  1. … 11 more files in changeset.
DEN-12106: Heparin Delivery Active Text Does Not Appear

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -3
    • +3
    ./connection/PreTreatmentConnectionStack.qml
    • -5
    • +5
    ./connection/PreTreatmentUltrafiltration.qml
  1. … 574 more files in changeset.
UI-BUG-S62: UI-BUG-S62-1

Addressed code reviews.

  1. … 5 more files in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 13 more files in changeset.
Merge branch 'UI_BUG_S62' into develop

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
DEN-11825: BP/HR screen has no instruction for user to read from bluetooth cuff or enter information

- Center aligned the notification

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 1 more file in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
Merge branch 'UI_BUG_S62' into develop

    • -0
    • +9
    ./connection/PreTreatmentConnectionStack.qml
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

    • -5
    • +6
    ./connection/PreTreatmentConnectionStack.qml
  1. … 3 more files in changeset.
DEN-11825: BP/HR screen has no instruction

- Added the suggested on line instruction for the user.

    • -0
    • +8
    ./connection/PreTreatmentConnectionStack.qml
DEN-8145: Issues with UI Create Treatment screen

- Restricted the Patient ID entry to be max 20 characters and has to have at least 1 alphanumeric character without space, regarding the conversation with KeborMed.

  1. … 2 more files in changeset.
Merge branch 'UI_BUG_S62' into develop

    • -2
    • +2
    ./connection/PreTreatmentConnectionStack.qml
  1. … 3 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

    • -2
    • +2
    ./connection/PreTreatmentConnectionStack.qml
  1. … 15 more files in changeset.
Code Cleanup

- DEN-11807: Treatment Confirmation - Rinseback Rate displays "min" instead of "mL/min"

- Renamed the Keyboard component name to VirtualKeyboard to resolve the InputPanel's Keyboard object/type conflict (There was no issue is was only a name conflict.)

- Preventing the Patient Id from being empty space string, and disabling the confirm in that case.

- Mode the create_update_folder.sh to the Scripts/local repository's folder.

- Added Qt logo and Diality Qt Colored log.

- Updated the alarmMapping.sh script to always checkout to common/staging and also print out the log on console.

-

  1. … 12 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

    • -309
    • +197
    ./create/PreTreatmentCreate.qml
  1. … 8 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10205: BCuff [Missing Feature]: Try to reconnect to the previously paired

Bluetooth Code improvements:

- stopping the agent from scanning if user selects a device to connect.

- Rewording of the parred device not found

- Added "ZERO" check for the address

- Added real-time Bluetooth status on top of the main screen.

- Not The UI will always updated but the TreatmentLog still needs UI to be in a proper screen/state to log the Vitals measurements.

Other Requested modifications:

Chris.Y:

- Added back for the pre-treatment BP/HR screen to the Ultrafiltration screen.

Eliza.P:

- Added notification for the Heparin Dispensing:"Heparin Active"

- Changed the button text to "OFF" when the heparin is off.

- Changed the button text to "COMPLETE" when the heparin is complete.

- Shortened the notification text and removed the "Heparin ... " from the beginning.

- Won't show the off notification since the button text changed to imply that.

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 10 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 10 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

    • -4
    • +54
    ./create/PreTreatmentCreateStack.qml
  1. … 8 more files in changeset.
DEN-9637: Update Design Documents

- Added the ability to enable/disable the vitals measurement logs [it's still unknown that how to stop in case of the treatment ended by alarm or system fault.]

- Added the Battery Level of the selected device.

    • -1
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 18 more files in changeset.