Merge branch 'staging' into DEN-15363-fw-handle-bld-sensor-communication-errors

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15314: Added stall check to syringe prime operation. Extended dialysate bypass prime timeout and added a little histeresis to exit criteria for bypass prime state.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15347: Addressed issue recovering from venous occlusion alarm 208.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-15317: Fixed issue with failure to populate pressure limit state in the pressure data broadcast msg x9.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-15359: Misc fixes to art/ven pressure limit changes.

  1. … 1 more file in changeset.
DEN-15359: Fix s/w fault bug. Change so that dialysate flow change will also kick pressure limit s.m. back to stabilization state.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15396-2-alarm-42-hd-fault-dpo-rotor-encountered-in-blood-prime-dvt-001' into staging

DEN-15359: Addressed code review comment.

DEN-15396 Code review comment, use macro.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15359: handle blood flow rate change in middle of stabilization state.

DEN-15359: broadcast pressure data immediately after window change by user.

DEN-15359: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-15359: finished arterial/venous pressure limits window update.

  1. … 3 more files in changeset.
DEN-15359: Additional implementation of arterial/venous pressure limit windows.

  1. … 3 more files in changeset.
DEN-15359: Added treatment parameter for asymmetric on venous pressure alarm limits.

  1. … 5 more files in changeset.
DEN-15359: Addressed DEN-15381.

  1. … 1 more file in changeset.
DEN-15359: Addressed DEN-15284.

DEN-15359: Updated tx params to change art/ven pressure limit params from min/max to window.

  1. … 4 more files in changeset.
DEN-13865: Addressed code review comment.

DEN-15396 Use greater of 10sec or 10 Rotor turn times for minimum persistent error time.

DEN-13865: Implemented alarm system feature to prevent alarm re-trigger per new properties. Addressed DEN-15295.

  1. … 3 more files in changeset.
DEN-15363 Reuse alarm ID 63 for Blood Leak communications errors

DEN-15396-2: Fixed rotor speed chk persistence function to handle negative motor speeds.