PresOccl.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-334 updated the code to initialize the variables in between the treatments without a need to power cycle the unit

  1. … 9 more files in changeset.
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-334_Alarm 127 Low Venous Pressure Not Triggered When Back TO Back TX IS Run Without Power Cycle
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

DENBUG-328 updated the min venous wide window range code

  1. … 2 more files in changeset.
DENBUG-328 added min venous limit from the institutional record

  1. … 2 more files in changeset.
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
HD-DENBUG-285_Alarm 141 Air Trap Fill Timeout
HD-DENBUG-285_Alarm 141 Air Trap Fill Timeout
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Delete the #define if not using it anymore.

Delete the #define if not using it anymore.

DENBUG-285 updated the code per the latest requests: 1. Reverted the VBV in airtrap. 2. Changed blood prime to 250mlPerMin and 3. Changed the min venous pressure to 0 mmHg

  1. … 1 more file in changeset.
HD-DENBUG-285_Alarm 141 Air Trap Fill Timeout
HD-DENBUG-285_Alarm 141 Air Trap Fill Timeout