dd-firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Remove extra paranthesis.

Remove extra paranthesis.

Remove empty lines

Remove empty lines

Mention in comment this is equivalent to Qd:600 minus 50ms switching period to stabilize the valve closure.

Mention in comment this is equivalent to Qd:600 minus 50ms switching period to stabilize the valve closure.

Update the comment for Spent chamber fill, max dialysate flow rate (Qd:600) used for fast filling.

Update the comment for Spent chamber fill, max dialysate flow rate (Qd:600) used for fast filling.

Same question as above: should we alarm if the retries failed?

Same question as above: should we alarm if the retries failed?

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

This else is not needed because isFlashAddressValid is set to False at the top of the function (Optional to change).

If a retry goes wrong will you trigger an alarm?

If a retry goes wrong will you trigger an alarm?

Any fault alarm here?

Any fault alarm here?

Have these variables been initialized correctly?

Have these variables been initialized correctly?

Remove extra blank line.

Remove extra blank line.

We generally don't add { } to switch cases unless they require local variable declarations.

We generally don't add { } to switch cases unless they require local variable declarations.

If we're not including a timestamp, can we just remove this line of code?

If we're not including a timestamp, can we just remove this line of code?

Why commented out?

Why commented out?

0.0F(float type).

0.0F(float type).

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

I meant 'handleBicarbChamberCartridgeFillWaterEndState function stops the inlet water to cartridge'.

Delete it.

Delete it.

Remove blank spaces after =.

Remove blank spaces after =.

To be deleted this line?

To be deleted this line?

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as w...

Add a comment here to wait for BC switching to go Idle state and transition to spent chamber fill state. Also, add a space after (). Apply the same comment in the next Dialysate delivery state as well.

I guess we can remove the "balancingChamberFillInProgress" check here.

I guess we can remove the "balancingChamberFillInProgress" check here.

Remove the TODO comment.

Remove the TODO comment.

align the comment.

align the comment.

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Vendor - TEL - Sameer PoyilWhat's the value you are recommending for?

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pre...

Remove values in the comment, as they are changing during development (200ms persistence, 4 PSI low pressure, 17 PSI high pressure etc.,), instead use defined persistence, low or high threshold pressure etc.,

decays to threshold level (4 PSI).

decays to threshold level (4 PSI).

remove all commented code.

remove all commented code.

Update comment. remove TODO if no more applicable.

Update comment. remove TODO if no more applicable.

FillWaterEndState?

FillWaterEndState?