It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD...

It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD or HDF type). Then when we get this response msg from UI, we will compare the type to the user selected modality (from TxParams) to see if it matches.

Add doxygen comments to the right of each structure field.

Add doxygen comments to the right of each structure field.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

Add blank line before return statement.

Add blank line before return statement.

Keep this blank line above group header.

Keep this blank line above group header.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Make sure UI s/w is looking for this message on the broadcast channel or change this channel to the TD=>UI channel if UI is looking for this message on that channel.

Make sure UI s/w is looking for this message on the broadcast channel or change this channel to the TD=>UI channel if UI is looking for this message on that channel.

LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
LEAHI-TD-FIRMWARE-LDT-3215_Measured venous pressure decreases to atmosphere if H12 air pump is dead headed
Maintain alphabetical order.

Maintain alphabetical order.

Eliza says we do allow multiple iso UF sessions in a treatment. So do we still need this else if, but looking at <= iso UF elapsed session time?

Eliza says we do allow multiple iso UF sessions in a treatment. So do we still need this else if, but looking at <= iso UF elapsed session time?

What about HDF?

What about HDF?

Once UI software sends the volume and/or duration change request and once we change them, the UI software will send another request to confirm the displayed settings to the user and that is what is...

Once UI software sends the volume and/or duration change request and once we change them, the UI software will send another request to confirm the displayed settings to the user and that is what is being handled in this function.

return result;

return result;

Keep local result and set to TRUE or FALSE with a ternary.

Keep local result and set to TRUE or FALSE with a ternary.

Which setting does this message confirm (duration or volume)?

Which setting does this message confirm (duration or volume)?

Update alarms and Input/Output sections.

Update alarms and Input/Output sections.

Also, add option individual parameter override for substitutions pump rate as well.

Also, add option individual parameter override for substitutions pump rate as well.

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

Space after parenthesis. Please add the same in else portion as well (when treatment is in progress, update the rate).

fixed.

fixed.

fixed

fixed

fixed

fixed

fixed

fixed

added

added

Could you update the Message ID to latest? 0XA9 is already used.

Could you update the Message ID to latest? 0XA9 is already used.

Align comment.

Align comment.

where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?